Modeling treatment couches in the Pinnacle treatment planning system: Especially important for arc therapy
- Ironwood Cancer and Research Center, Chandler, AZ (United States)
- Department of Radiation Oncology, University of Mississippi Medical Center, Jackson, MS (United States)
This study is to demonstrate the importance and a method of properly modeling the treatment couch for dose calculation in patient treatment using arc therapy. The 2 treatment couch tops—Aktina AK550 and Elekta iBEAM evo—of Elekta LINACs were scanned using Philips Brilliance Big Bore CT Simulator. Various parts of the couch tops were contoured, and their densities were measured and recorded on the Pinnacle treatment planning system (TPS) using the established computed tomography density table. These contours were saved as organ models to be placed beneath the patient during planning. Relative attenuation measurements were performed following procedures outlined by TG-176 as well as absolute dose comparison of static fields of 10 × 10 cm{sup 2} that were delivered through the couch tops with that calculated in the TPS with the couch models. A total of 10 random arc therapy treatment plans (5 volumetric-modulated arc therapy [VMAT] and 5 stereotactic body radiation therapy [SBRT]), using 24 beams, were selected for this study. All selected plans were calculated with and without couch modeling. Each beam was evaluated using the Delta{sup 4} dosimetry system (Delta{sup 4}). The Student t-test was used to determine statistical significance. Independent reviews were exploited as per the Imaging and Radiation Oncology Core head and neck credentialing phantom. The selected plans were calculated on the actual patient anatomies with and without couch modeling to determine potential clinical effects. Large relative beam attenuations were noted dependent on which part of the couch top beams were passing through. Substantial improvements were also noted for static fields both calculated with the TPS and delivered physically when the couch models were included in the calculation. A statistically significant increase in agreement was noted for dose difference, distance to agreement, and γ-analysis with the Delta{sup 4} on VMAT and SBRT plans. A credentialing review showed improvement in treatment delivery after couch modeling with both thermoluminescent dosimeter doses and film analysis. Furthermore, analysis of treatment plans with and without using the couch model showed a statistically significant reduction in planning target volume coverage and increase in skin dose. In conclusion, ignoring the treatment couch, a common practice when generating a patient treatment plan, can overestimate the dose delivered especially for arc therapy. This work shows that explicitly modeling the couch during planning can meaningfully improve the agreement between calculated and measured dose distributions. Because of this project, we have implemented the couch models clinically across all treatment plans.
- OSTI ID:
- 22577858
- Journal Information:
- Medical Dosimetry, Vol. 41, Issue 1; Other Information: Copyright (c) 2016 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.; Country of input: International Atomic Energy Agency (IAEA); ISSN 0958-3947
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-E-T-115: Analysis of Patient Specific QA for VMAT by Disease Site and Planning-Delivery System Using the ScandiDos Delta4 Phantom
SU-F-T-298: The Impact of Modeling the Treatment Couch On Patient Specific VMAT QA
Related Subjects
62 RADIOLOGY AND NUCLEAR MEDICINE
ANATOMY
ATTENUATION
BEAMS
BIOMEDICAL RADIOGRAPHY
COMPUTERIZED TOMOGRAPHY
FILM DOSIMETRY
HEAD
LINEAR ACCELERATORS
NECK
PATIENTS
PHANTOMS
PLANNING
RADIATION DOSE DISTRIBUTIONS
RADIATION DOSES
RADIOTHERAPY
SIMULATION
SIMULATORS
SKIN
THERMOLUMINESCENT DOSEMETERS
VALIDATION