SU-F-T-298: The Impact of Modeling the Treatment Couch On Patient Specific VMAT QA
- University of Iowa Hospitals & Clinics, Iowa City, IA (United States)
Purpose: The aim of this work is to quantify the impact of modeling the treatment couch on the passing rate of ion chamber measurements during VMAT quality assurance. Methods: For the initial characterization, attenuation and surface dose measurements were performed following the guidelines of TG-176 for the Civco Universal couch top using an Elekta VersaHD accelerator at an energy of 6 MV. A simulation CT was performed to aid in the creation of contours for representing the shape and size of the couch top in the treatment planning system (TPS). A uniform value of density for the couch wall was determined by comparing the ratios of ion chamber measurements made in a 30×30×11 cm3 water phantom with the TPS dose values of a plan with the same geometry. At our institution, patient specific quality assurance is performed using a Sun Nuclear ArcCheck with a multi-plug for chamber measurements, a 0.125cc PTW TN31010 chamber, and a Sun Nuclear 1010 electrometer. Ten VMAT plans were transferred into the phantom geometry created in the TPS with two settings: with and without the couch. The chamber measurements were compared to both treatment plans. Results: A maximum attenuation of 3.6% was observed when the gantry angle was set to 120 and 240 degrees, passing obliquely through the couch. A uniform density of 0.6 g/cm3 for the couch wall was determined in the TPS by comparison with measured data. The VMAT ion chamber measurement/plan ratios systematically improved by 1.79% ±0.53% for all patients when the couch was included in the calculation. Conclusion: The attenuation and surface dose changes produced by the Civco couch can generate observable dose difference in VMAT plans. Including a couch model in the phantom plan used for patient specific VMAT QA can improve the ionization chamber agreement by up to ∼2%.
- OSTI ID:
- 22648906
- Journal Information:
- Medical Physics, Vol. 43, Issue 6; Other Information: (c) 2016 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
VMAT QA: Measurement-guided 4D dose reconstruction on a patient
SU-E-T-650: Quantification and Modeling of the Dosimetric Impact of the IBEAM Evo Treatment Couchtop EP (Elekta) in VMAT