SU-E-T-624: Quantitative Evaluation of 2D Versus 3D Dosimetry for Stereotactic Volumetric Modulated Arc Delivery Using COMPASS
- Medanta The Medicity, Gurgaon, Haryana (India)
- Sri Siddhivinayak Ganapathi Cancer hospital, Miraj, Maharastra (India)
- St Johns Medical College, Bangalore, Karnataka (India)
- Apollo Super Speciality Hospital, Chennai, Tamil Nadu (India)
Purpose: The purpose of this study was to evaluate quantitatively 2D versus 3D dosimetry for stereotactic volumetric modulated arc delivery using COMPASS with 2D array. Methods: Twenty-five patients CT images and RT structures of different sites like brain, head and neck, thorax, abdomen and spine were taken from Multiplan planning system for this study. All these patients underwent radical stereotactic treatment in Cyberknife. For each patient, linac based VMAT stereotactic plans were generated in Monaco TPS v 3.1 using Elekta Beam Modulator MLC. Dose prescription was in the range of 5-20Gy/fraction.TPS calculated VMAT plan delivery accuracy was quantitatively evaluated with COMPASS measured dose and calculated dose based on DVH metrics. In order to ascertain the potential of COMPASS 3D dosimetry for stereotactic plan delivery, 2D fluence verification was performed with MatriXX using Multicube. Results: For each site, D{sub 9} {sub 5} was achieved with 100% of prescription dose with maximum 0.05SD. Conformity index (CI) was observed closer to 1.15 in all cases. Maximum deviation of 2.62 % was observed for D{sub 9} {sub 5} when compared TPS versus COMPASS measured. Considerable deviations were observed in head and neck cases compare to other sites. The maximum mean and standard deviation for D{sub 9} {sub 5}, average target dose and average gamma were -0.78±1.72, -1.10±1.373 and 0.39±0.086 respectively. Numbers of pixels passing 2D fluence verification were observed as a mean of 99.36% ±0.455 SD with 3% dose difference and 3mm DTA. For critical organs in head and neck cases, significant dose differences were observed in 3D dosimetry while the target doses were matched well within limit in both 2D and 3D dosimetry. Conclusion: The quantitative evaluations of 2D versus 3D dosimetry for stereotactic volumetric modulated plans showed the potential of highlighting the delivery errors. This study reveals that COMPASS 3D dosimetry is an effective tool for patient specific quality assurance compared to 2D fluence verification.
- OSTI ID:
- 22402229
- Journal Information:
- Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 41; ISSN 0094-2405; ISSN MPHYA6
- Country of Publication:
- United States
- Language:
- English
Similar Records
SU-E-T-77: Comparison of 2D and 3D Gamma Analysis in Patient-Specific QA for Prostate VMAT Plans
SU-E-T-115: Analysis of Patient Specific QA for VMAT by Disease Site and Planning-Delivery System Using the ScandiDos Delta4 Phantom
Validation of a method for in vivo 3D dose reconstruction for IMRT and VMAT treatments using on-treatment EPID images and a model-based forward-calculation algorithm
Journal Article
·
Sun Jun 01 00:00:00 EDT 2014
· Medical Physics
·
OSTI ID:22339844
SU-E-T-115: Analysis of Patient Specific QA for VMAT by Disease Site and Planning-Delivery System Using the ScandiDos Delta4 Phantom
Journal Article
·
Mon Jun 15 00:00:00 EDT 2015
· Medical Physics
·
OSTI ID:22545241
Validation of a method for in vivo 3D dose reconstruction for IMRT and VMAT treatments using on-treatment EPID images and a model-based forward-calculation algorithm
Journal Article
·
Mon Dec 14 23:00:00 EST 2015
· Medical Physics
·
OSTI ID:22482428