Validation of a method for in vivo 3D dose reconstruction for IMRT and VMAT treatments using on-treatment EPID images and a model-based forward-calculation algorithm

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4935199· OSTI ID:22482428
 [1];  [2];  [3];  [4];  [5]
  1. Medical Physics Department, CancerCare Manitoba, 675 McDermot Avenue, Winnipeg, Manitoba R3E 0V9 (Canada)
  2. Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba R3T 2N2, Canada and Medical Physics Department, CancerCare Manitoba, 675 McDermot Avenue, Winnipeg, Manitoba R3E 0V9 (Canada)
  3. Medical Physics Department, Nova Scotia Cancer Centre, 5820 University Avenue, Halifax, Nova Scotia B3H 1V7 (Canada)
  4. School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW 2308 (Australia)
  5. Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba R3T 2N2 (Canada)
Purpose: Radiation treatments are trending toward delivering higher doses per fraction under stereotactic radiosurgery and hypofractionated treatment regimens. There is a need for accurate 3D in vivo patient dose verification using electronic portal imaging device (EPID) measurements. This work presents a model-based technique to compute full three-dimensional patient dose reconstructed from on-treatment EPID portal images (i.e., transmission images). Methods: EPID dose is converted to incident fluence entering the patient using a series of steps which include converting measured EPID dose to fluence at the detector plane and then back-projecting the primary source component of the EPID fluence upstream of the patient. Incident fluence is then recombined with predicted extra-focal fluence and used to calculate 3D patient dose via a collapsed-cone convolution method. This method is implemented in an iterative manner, although in practice it provides accurate results in a single iteration. The robustness of the dose reconstruction technique is demonstrated with several simple slab phantom and nine anthropomorphic phantom cases. Prostate, head and neck, and lung treatments are all included as well as a range of delivery techniques including VMAT and dynamic intensity modulated radiation therapy (IMRT). Results: Results indicate that the patient dose reconstruction algorithm compares well with treatment planning system computed doses for controlled test situations. For simple phantom and square field tests, agreement was excellent with a 2%/2 mm 3D chi pass rate ≥98.9%. On anthropomorphic phantoms, the 2%/2 mm 3D chi pass rates ranged from 79.9% to 99.9% in the planning target volume (PTV) region and 96.5% to 100% in the low dose region (>20% of prescription, excluding PTV and skin build-up region). Conclusions: An algorithm to reconstruct delivered patient 3D doses from EPID exit dosimetry measurements was presented. The method was applied to phantom and patient data sets, as well as for dynamic IMRT and VMAT delivery techniques. Results indicate that the EPID dose reconstruction algorithm presented in this work is suitable for clinical implementation.
OSTI ID:
22482428
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 12 Vol. 42; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

Sci—Thur AM: YIS - 06: An EPID-based 3D patient dose verification method for SBRT-VMAT delivery
Journal Article · Fri Aug 15 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22409502

An in vivo dose verification method for SBRT–VMAT delivery using the EPID
Journal Article · Mon Dec 14 23:00:00 EST 2015 · Medical Physics · OSTI ID:22482429

SIFT: A method to verify the IMRT fluence delivered during patient treatment using an electronic portal imaging device
Journal Article · Sun Oct 31 23:00:00 EST 2004 · International Journal of Radiation Oncology, Biology and Physics · OSTI ID:20630952