The development of in-situ calibration method for divertor IR thermography in ITER
Journal Article
·
· AIP Conference Proceedings
- Japan Atomic Energy Agency (Japan)
For the development of the calibration method of the emissivity in IR light on the divertor plate in ITER divertor IR thermography system, the laboratory experiments have been performed by using IR instruments. The calibration of the IR camera was performed by the plane black body in the temperature of 100–600 degC. The radiances of the tungsten heated by 280 degC were measured by the IR camera without filter (2.5–5.1 μm) and with filter (2.95 μm, 4.67 μm). The preliminary data of the scattered light of the laser of 3.34 μm that injected into the tungsten were acquired.
- OSTI ID:
- 22308283
- Journal Information:
- AIP Conference Proceedings, Journal Name: AIP Conference Proceedings Journal Issue: 1 Vol. 1612; ISSN APCPCS; ISSN 0094-243X
- Country of Publication:
- United States
- Language:
- English
Similar Records
Divertor IR thermography on Alcator C-Mod
Infrared thermography of the DIII-D divertor targets
Infrared thermography of the DIII-D divertor targets
Journal Article
·
Fri Oct 15 00:00:00 EDT 2010
· Review of Scientific Instruments
·
OSTI ID:22055889
Infrared thermography of the DIII-D divertor targets
Journal Article
·
Mon Aug 01 00:00:00 EDT 1988
· Rev. Sci. Instrum.; (United States)
·
OSTI ID:7123460
Infrared thermography of the DIII-D divertor targets
Conference
·
Sun Jan 31 23:00:00 EST 1988
·
OSTI ID:6986067