On determining dose rate constants spectroscopically

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4770284· OSTI ID:22099171
;  [1]
  1. Carleton Laboratory for Radiotherapy Physics, Carleton University, Ottawa K1S 5B6 (Canada)
Purpose: To investigate several aspects of the Chen and Nath spectroscopic method of determining the dose rate constants of {sup 125}I and {sup 103}Pd seeds [Z. Chen and R. Nath, Phys. Med. Biol. 55, 6089-6104 (2010)] including the accuracy of using a line or dual-point source approximation as done in their method, and the accuracy of ignoring the effects of the scattered photons in the spectra. Additionally, the authors investigate the accuracy of the literature's many different spectra for bare, i.e., unencapsulated {sup 125}I and {sup 103}Pd sources. Methods: Spectra generated by 14 {sup 125}I and 6 {sup 103}Pd seeds were calculated in vacuo at 10 cm from the source in a 2.7 Multiplication-Sign 2.7 Multiplication-Sign 0.05 cm{sup 3} voxel using the EGSnrc BrachyDose Monte Carlo code. Calculated spectra used the initial photon spectra recommended by AAPM's TG-43U1 and NCRP (National Council of Radiation Protection and Measurements) Report 58 for the {sup 125}I seeds, or TG-43U1 and NNDC(2000) (National Nuclear Data Center, 2000) for {sup 103}Pd seeds. The emitted spectra were treated as coming from a line or dual-point source in a Monte Carlo simulation to calculate the dose rate constant. The TG-43U1 definition of the dose rate constant was used. These calculations were performed using the full spectrum including scattered photons or using only the main peaks in the spectrum as done experimentally. Statistical uncertainties on the air kerma/history and the dose rate/history were Less-Than-Or-Slanted-Equal-To 0.2%. The dose rate constants were also calculated using Monte Carlo simulations of the full seed model. Results: The ratio of the intensity of the 31 keV line relative to that of the main peak in {sup 125}I spectra is, on average, 6.8% higher when calculated with the NCRP Report 58 initial spectrum vs that calculated with TG-43U1 initial spectrum. The {sup 103}Pd spectra exhibit an average 6.2% decrease in the 22.9 keV line relative to the main peak when calculated with the TG-43U1 rather than the NNDC(2000) initial spectrum. The measured values from three different investigations are in much better agreement with the calculations using the NCRP Report 58 and NNDC(2000) initial spectra with average discrepancies of 0.9% and 1.7% for the {sup 125}I and {sup 103}Pd seeds, respectively. However, there are no differences in the calculated TG-43U1 brachytherapy parameters using either initial spectrum in both cases. Similarly, there were no differences outside the statistical uncertainties of 0.1% or 0.2%, in the average energy, air kerma/history, dose rate/history, and dose rate constant when calculated using either the full photon spectrum or the main-peaks-only spectrum. Conclusions: Our calculated dose rate constants based on using the calculated on-axis spectrum and a line or dual-point source model are in excellent agreement (0.5% on average) with the values of Chen and Nath, verifying the accuracy of their more approximate method of going from the spectrum to the dose rate constant. However, the dose rate constants based on full seed models differ by between +4.6% and -1.5% from those based on the line or dual-point source approximations. These results suggest that the main value of spectroscopic measurements is to verify full Monte Carlo models of the seeds by comparison to the calculated spectra.
OSTI ID:
22099171
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 1 Vol. 40; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English