Optical diagnostic suite (schlieren, interferometry, and grid image refractometry) on OMEGA EP using a 10-ps, 263-nm probe beam

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4733739· OSTI ID:22093962
; ; ; ; ; ; ; ; ;  [1]; ;  [2]
  1. Laboratory for Laser Energetics, University of Rochester, 250 E. River Rd., Rochester, New York 14616 (United States)
  2. Lawrence Livermore National Laboratory, University of California, P. O. Box 808, Livermore, California 94551 (United States)

A 10-ps, 263-nm (4{omega}) laser is being built to probe plasmas produced on the OMEGA EP [J. H. Kelly, L. J. Waxer, V. Bagnoud, I. A. Begishev, J. Bromage, B. E. Kruschwitz, T. E. Kessler, S. J. Loucks, D. N. Maywar, R. L. McCrory et al., J. Phys. IV France 133, 75-80 (2006)]. A suite of optical diagnostics (schlieren, interferometry, and grid image refractometry) has been designed to diagnose and characterize a wide variety of plasmas. Light scattered by the probe beam is collected by an f/4 catadioptric telescope and a transport system is designed to image with a near-diffraction-limited resolution ({approx}1 -{mu}m full width at half maximum) over a 5-mm field of view to a diagnostic table. The transport system provides a contrast greater than 1 : 10{sup 4} with respect to all wavelengths outside of the 263 {+-} 2 nm measurement range.

OSTI ID:
22093962
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 10 Vol. 83; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Optical diagnostic suite (schlieren, interferometry, and grid image refractometry) on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article · Mon Oct 01 00:00:00 EDT 2012 · Review of Scientific Instruments · OSTI ID:1053477

Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article · Mon Aug 11 00:00:00 EDT 2014 · Review of Scientific Instruments · OSTI ID:1156677

Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article · Fri Nov 14 23:00:00 EST 2014 · Review of Scientific Instruments · OSTI ID:22308650