Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article
·
· Review of Scientific Instruments
- University of Rochester Laboratory for Laser Energetics, , Rochester, New York 14623, USA; Laboratory for Laser Energetics, University of Rochester, Rochester, NY
- University of Rochester Laboratory for Laser Energetics, , Rochester, New York 14623, USA
A polarimetry diagnostic was built and characterized for magnetic-field measurements in laser-plasma experiments on the OMEGA EP laser. This diagnostic was built into the existing 4ω (263-nm) probe system that employs a 10-ps laser pulse collected with an f/4 imaging system. The diagnostic measures the rotation of the probe beam's polarization. The polarimeter uses a Wollaston prism to split the probe beam into orthogonal polarization components. Spatially localized intensity variations between images indicate polarization rotation. Magnetic fields can be calculated by combining the polarimetry data with the measured plasma density profile obtained from angular filter refractometry.
- Research Organization:
- Univ. of Rochester, NY (United States). Lab. for Laser Energetics
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- NA0001944
- OSTI ID:
- 1156677
- Report Number(s):
- DOE/NA/0001944-1199; 2014-21; 2166
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 11 Vol. 85; ISSN 0034-6748
- Publisher:
- American Institute of Physics (AIP)
- Country of Publication:
- United States
- Language:
- English
Similar Records
Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam
Optical diagnostic suite (schlieren, interferometry, and grid image refractometry) on OMEGA EP using a 10-ps, 263-nm probe beam
Implementation of a Wollaston interferometry diagnostic on OMEGA EP
Journal Article
·
Fri Nov 14 23:00:00 EST 2014
· Review of Scientific Instruments
·
OSTI ID:22308650
Optical diagnostic suite (schlieren, interferometry, and grid image refractometry) on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article
·
Mon Oct 01 00:00:00 EDT 2012
· Review of Scientific Instruments
·
OSTI ID:1053477
Implementation of a Wollaston interferometry diagnostic on OMEGA EP
Journal Article
·
Thu Aug 02 00:00:00 EDT 2018
· Review of Scientific Instruments
·
OSTI ID:1468612