Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4889908· OSTI ID:1156677
 [1];  [2];  [2];  [2];  [2];  [2]
  1. University of Rochester Laboratory for Laser Energetics, , Rochester, New York 14623, USA; Laboratory for Laser Energetics, University of Rochester, Rochester, NY
  2. University of Rochester Laboratory for Laser Energetics, , Rochester, New York 14623, USA

A polarimetry diagnostic was built and characterized for magnetic-field measurements in laser-plasma experiments on the OMEGA EP laser. This diagnostic was built into the existing 4ω (263-nm) probe system that employs a 10-ps laser pulse collected with an f/4 imaging system. The diagnostic measures the rotation of the probe beam's polarization. The polarimeter uses a Wollaston prism to split the probe beam into orthogonal polarization components. Spatially localized intensity variations between images indicate polarization rotation. Magnetic fields can be calculated by combining the polarimetry data with the measured plasma density profile obtained from angular filter refractometry.

Research Organization:
Univ. of Rochester, NY (United States). Lab. for Laser Energetics
Sponsoring Organization:
USDOE
DOE Contract Number:
NA0001944
OSTI ID:
1156677
Report Number(s):
DOE/NA/0001944-1199; 2014-21; 2166
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 11 Vol. 85; ISSN 0034-6748
Publisher:
American Institute of Physics (AIP)
Country of Publication:
United States
Language:
English

References (20)

Development of the indirect‐drive approach to inertial confinement fusion and the target physics basis for ignition and gain journal November 1995
Magnetic-field generation in laser fusion and hot-electron transport journal August 1986
Optical diagnostic suite (schlieren, interferometry, and grid image refractometry) on OMEGA EP using a 10-ps, 263-nm probe beam journal October 2012
Measurements of electron density profiles using an angular filter refractometer journal May 2014
Magnetic reconnection in high-energy-density laser-produced plasmas journal May 2012
Proton deflectometry of a magnetic reconnection geometry journal April 2010
Fast Magnetic Reconnection in Laser-Produced Plasma Bubbles journal May 2011
Bidirectional jet formation during driven magnetic reconnection in two-beam laser–plasma interactions journal September 2008
High-Energy Petawatt Capability for the Omega Laser journal January 2005
A magnetic reconnection X-line extending more than 390 Earth radii in the solar wind journal January 2006
Faraday Effect in a Plasma journal December 1965
Hydrogen Phases on the Surfaces of a Strongly Magnetized Neutron Star journal December 1997
Fusion Yield Enhancement in Magnetized Laser-Driven Implosions journal July 2011
Modelling loop-top X-ray source and reconnection outflows in solar flares with intense lasers journal October 2010
A loop-top hard X-ray source in a compact solar flare as evidence for magnetic reconnection journal October 1994
Quenching of the Nonlocal Electron Heat Transport by Large External Magnetic Fields in a Laser-Produced Plasma Measured with Imaging Thomson Scattering journal March 2007
Review on spontaneous magnetic fields in laser-produced plasmas: Phenomena and measurements journal December 1991
Magnetic Reconnection and Plasma Dynamics in Two-Beam Laser-Solid Interactions journal December 2006
Thomson scattering from laser plasmas journal May 1999
Observation of Megagauss-Field Topology Changes due to Magnetic Reconnection in Laser-Produced Plasmas journal August 2007

Similar Records

Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article · Fri Nov 14 23:00:00 EST 2014 · Review of Scientific Instruments · OSTI ID:22308650

Optical diagnostic suite (schlieren, interferometry, and grid image refractometry) on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article · Mon Oct 01 00:00:00 EDT 2012 · Review of Scientific Instruments · OSTI ID:1053477

Implementation of a Wollaston interferometry diagnostic on OMEGA EP
Journal Article · Thu Aug 02 00:00:00 EDT 2018 · Review of Scientific Instruments · OSTI ID:1468612

Related Subjects