Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.4889908· OSTI ID:22308650
; ; ; ;  [1]
  1. Laboratory for Laser Energetics, University of Rochester, Rochester, New York 14623 (United States)

A polarimetry diagnostic was built and characterized for magnetic-field measurements in laser-plasma experiments on the OMEGA EP laser. This diagnostic was built into the existing 4ω (263-nm) probe system that employs a 10-ps laser pulse collected with an f/4 imaging system. The diagnostic measures the rotation of the probe beam's polarization. The polarimeter uses a Wollaston prism to split the probe beam into orthogonal polarization components. Spatially localized intensity variations between images indicate polarization rotation. Magnetic fields can be calculated by combining the polarimetry data with the measured plasma density profile obtained from angular filter refractometry.

OSTI ID:
22308650
Journal Information:
Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 11 Vol. 85; ISSN 0034-6748; ISSN RSINAK
Country of Publication:
United States
Language:
English

Similar Records

Polarimetry diagnostic on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article · Mon Aug 11 00:00:00 EDT 2014 · Review of Scientific Instruments · OSTI ID:1156677

Optical diagnostic suite (schlieren, interferometry, and grid image refractometry) on OMEGA EP using a 10-ps, 263-nm probe beam
Journal Article · Mon Oct 01 00:00:00 EDT 2012 · Review of Scientific Instruments · OSTI ID:1053477

Implementation of a Wollaston interferometry diagnostic on OMEGA EP