2018 Kinemetrics Q330M+ Digitizer Evaluation

Technical Report ·
DOI:https://doi.org/10.2172/1762654· OSTI ID:1762654
 [1];  [1]
  1. Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
Sandia National Laboratories has tested and evaluated two Kinemetrics Q330M+ digitizers. The digitizers are intended to record sensor output for seismic and infrasound monitoring applications. Notable improvements to the Q330M+ include the support for transmission and authentication of CD1.1 data, integration of analog and digital weather stations, support for multiple gain amplification levels, and the use of a webpage for status and configuration of the digitizer. The purpose of this digitizer evaluation is to measure the performance characteristics in such areas as power consumption, input impedance, sensitivity, full scale, self-noise, dynamic range, system noise, response, passband, and timing. The digitizers are being evaluated for potential use in the International Monitoring System (IMS) of the Comprehensive Nuclear Test- Ban-Treaty Organization (CTBTO).
Research Organization:
Sandia National Laboratories (SNL-NM), Albuquerque, NM (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA); Comprehensive Nuclear Test-Ban-Treaty Organization (CTBTO)
DOE Contract Number:
AC04-94AL85000; NA0003525
OSTI ID:
1762654
Report Number(s):
SAND--2019-6501; 676255
Country of Publication:
United States
Language:
English

Similar Records

2018 Nanometrics Centaur Digitizer Evaluation
Technical Report · Mon Oct 01 00:00:00 EDT 2018 · OSTI ID:1481600

2017 Guralp Affinity Digitizer Evaluation.
Technical Report · Wed Feb 28 23:00:00 EST 2018 · OSTI ID:1425328

CEA SMAD 2016 Digitizer Evaluation.
Technical Report · Tue Aug 01 00:00:00 EDT 2017 · OSTI ID:1374793