Comparison of gamma-ray spectral analysis methods for thick-walled UF6 cylinders

Journal Article · · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Three data analysis methods were compared for gamma ray spectra collected by planar and coaxial high-purity germanium detectors (HPGe) on thick-walled uranium hexafluoride cylinders. The cylinder contents range from natural to low-enriched uranium (4.95% 235U). Each spectrum was analyzed with the traditional 185.7 keV enrichment meter method, a 185.7/1001 keV ratio method, and the Fixed Energy Response Function Analysis with Multiple Efficiency (FRAM) software that utilizes low- and high-energy gamma rays and X-rays. When results for all cylinders were compared, the enrichment meter method and the 185.7/1001 keV peak ratio method consistently provided the most accurate data for all detector types. The error on the enrichment calculations analyzed with FRAM did not meet the international target value whereas those calculated with the enrichment meter and ratio method did. Enrichments calculated by FRAM were found to skew to higher values on natural uranium and to lower values on low-enriched uranium. FRAM calculated the least-precise enrichments; replicate analyses of single cylinders resulted in 235U values that varied by as much as 19%. FRAM does have the significant advantages in that it does not require the use of calibration cylinders, infinite thickness criteria are not necessary, and it can be applied to any HPGe detector. Finally, if calibration cylinders are available, the enrichment meter and 185.7/1001 peak ratio are the recommended methods as they are the most accurate and precise.
Research Organization:
Oak Ridge National Laboratory (ORNL), Oak Ridge, TN (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA)
Grant/Contract Number:
AC05-00OR22725
OSTI ID:
1649514
Alternate ID(s):
OSTI ID: 1635431
Journal Information:
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment, Journal Name: Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment Journal Issue: 1 Vol. 977; ISSN 0168-9002
Publisher:
ElsevierCopyright Statement
Country of Publication:
United States
Language:
English

References (4)

PC/FRAM: a code for the nondestructive measurement of the isotopic composition of actinides for safeguards applications journal October 1997
Study on the performance of different uranium isotopic codes used in nuclear safeguards activities journal April 2019
An unattended verification station for UF 6 cylinders: Field trial findings
  • Smith, L. E.; Miller, K. A.; McDonald, B. S.
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 874 https://doi.org/10.1016/j.nima.2017.08.003
journal December 2017
Measurements of UF 6 Cylinders with Portable Instruments journal February 1974

Similar Records

Heterogeneity effects on nondestructive assay measurements of enrichment in UF6 cylinders
Journal Article · Mon Nov 30 19:00:00 EST 2020 · ESARDA Bulletin · OSTI ID:1756280

Applications of HPGe-detected high energy gamma rays toward quantifying neutron emission rates and 234U enrichment in UF6 cylinders
Journal Article · Fri Apr 10 20:00:00 EDT 2020 · Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment · OSTI ID:1631249

FRAM isotopic analysis of uranium in thick-walled containers using high energy gamma rays and planar HPGe detectors.
Conference · Mon Dec 31 23:00:00 EST 2001 · OSTI ID:976166