Comparison of gamma-ray spectral analysis methods for thick-walled UF6 cylinders
- Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)
Three data analysis methods were compared for gamma ray spectra collected by planar and coaxial high-purity germanium detectors (HPGe) on thick-walled uranium hexafluoride cylinders. The cylinder contents range from natural to low-enriched uranium (4.95% 235U). Each spectrum was analyzed with the traditional 185.7 keV enrichment meter method, a 185.7/1001 keV ratio method, and the Fixed Energy Response Function Analysis with Multiple Efficiency (FRAM) software that utilizes low- and high-energy gamma rays and X-rays. When results for all cylinders were compared, the enrichment meter method and the 185.7/1001 keV peak ratio method consistently provided the most accurate data for all detector types. The error on the enrichment calculations analyzed with FRAM did not meet the international target value whereas those calculated with the enrichment meter and ratio method did. Enrichments calculated by FRAM were found to skew to higher values on natural uranium and to lower values on low-enriched uranium. FRAM calculated the least-precise enrichments; replicate analyses of single cylinders resulted in 235U values that varied by as much as 19%. FRAM does have the significant advantages in that it does not require the use of calibration cylinders, infinite thickness criteria are not necessary, and it can be applied to any HPGe detector. Finally, if calibration cylinders are available, the enrichment meter and 185.7/1001 peak ratio are the recommended methods as they are the most accurate and precise.
- Research Organization:
- Oak Ridge National Laboratory (ORNL), Oak Ridge, TN (United States)
- Sponsoring Organization:
- USDOE National Nuclear Security Administration (NNSA)
- Grant/Contract Number:
- AC05-00OR22725
- OSTI ID:
- 1649514
- Alternate ID(s):
- OSTI ID: 1635431
- Journal Information:
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment, Vol. 977, Issue 1; ISSN 0168-9002
- Publisher:
- ElsevierCopyright Statement
- Country of Publication:
- United States
- Language:
- English
Web of Science
Measurements of UF 6 Cylinders with Portable Instruments
|
journal | February 1974 |
Study on the performance of different uranium isotopic codes used in nuclear safeguards activities
|
journal | April 2019 |
An unattended verification station for UF cylinders: Field trial findings
|
journal | December 2017 |
PC/FRAM: a code for the nondestructive measurement of the isotopic composition of actinides for safeguards applications
|
journal | October 1997 |
Similar Records
Applications of HPGe-detected high energy gamma rays toward quantifying neutron emission rates and 234U enrichment in UF6 cylinders
FOUR PI CALIBRATION AND MODELING OF A BARE GERMANIUM DETECTOR IN A CYLINDRICAL FIELD SOURCE