Mechanical Design of a UHV Non-magnetic Kappa Diffractometer
A UHV non-magnetic kappa geometry diffractometer has been designed for the resonant soft X-ray scattering (RSXS) branch of the Intermediate Energy X-ray (IEX) beamline at the Advanced Photon Source (APS), Argonne National Laboratory. The kappa geometry allows for a large q-range and space for a sample environment (electric or magnetic fields). To avoid distorting the sample environment field, non-magnetic SmarAct, GmbH stages are used for the 4)-axis, X, Y, and Z linear stages. The ic and 0 axes are controlled by high-precision UHV compatible Physik Instrumente (PI) GmbH & Co. rotation stages. The design goal was to achieve a 50 gm sphere of confusion. To achieve this, iterative designs were developed using FEA and commercial stages were chosen that had minimal wobble, eccentricity, and flatness errors.
- Research Organization:
- Argonne National Laboratory (ANL)
- Sponsoring Organization:
- USDOE Office of Science - Office of Basic Energy Sciences
- DOE Contract Number:
- AC02-06CH11357
- OSTI ID:
- 1560668
- Country of Publication:
- United States
- Language:
- English
The intermediate energy X-ray beamline at the APS
|
journal | May 2014 |
Similar Records
First testing of the fast kappa diffractometers at National Synchrotron Light Source and European Synchrotron Radiation Facility
Surface And Interface Studies At APS Endstation 5ID-C
Diffractometer-Control Software For Bragg-Rod Measurements
Journal Article
·
Tue Jan 31 23:00:00 EST 1995
· Review of Scientific Instruments; (United States)
·
OSTI ID:6597251
Surface And Interface Studies At APS Endstation 5ID-C
Journal Article
·
Wed May 12 00:00:00 EDT 2004
· AIP Conference Proceedings
·
OSTI ID:20652832
Diffractometer-Control Software For Bragg-Rod Measurements
Journal Article
·
Wed May 12 00:00:00 EDT 2004
· AIP Conference Proceedings
·
OSTI ID:20652846