Quantitative measurements of cerebral blood flow in volume imaging PET scanners

Journal Article · · IEEE Transactions on Nuclear Science
DOI:https://doi.org/10.1109/23.467755· OSTI ID:148042
; ; ; ;  [1]
  1. Univ. of Pennsylvania, Philadelphia, PA (United States)
Quantitative measurements of Cerebral Blood Flow (CBF) are performed in a volume imaging PET Scanner by means of moderate activity infusions. In equilibrium infusions, activations are measured by scanning over 10 minutes with 16 minute activations. Typical measured whole brain CBF values are 37{+-}8 ml/min/100g, close to the value of 42 ml/min/100g reported by other groups using this method. For ramped infusions, scanning over 4 minutes with 5 minute activations results in whole brain CBFs of 49 {+-} 9 ml/min/100g, close to the Kety and Schmidt value of 50 ml/min/100g. Both equilibrium and ramped infusion methods have been used to study face and word memory in human subjects. Both methods were able to detect significant activations in regions implicated in human memory. The authors conclude that precise quantitation of regional CBF is achieved using both methods, and that ramped infusions also provide accurate measures of CBF. In addition a simplified protocol for ramped infusion studies has been developed. In this method the whole brain tissue time activity curve generated from dynamic scanning is replaced by an appropriately scaled camera coincidence countrate curve. The resulting whole brain CBF values are only 7% different from the dynamic scan and fit results. Regional CBFs (rCBF) may then be generated from the summed image (4.25 minutes) using a count density vs flow lookup table.
OSTI ID:
148042
Report Number(s):
CONF-941061--
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 4Pt1 Vol. 42; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

Positron imaging of intravenous O-15 labeled water for cerebral blood flow
Conference · Sat Dec 31 23:00:00 EST 1983 · J. Nucl. Med.; (United States) · OSTI ID:6843249

Regional cerebral blood flow abnormalities in Alzheimer's Disease
Conference · Wed May 01 00:00:00 EDT 1985 · J. Nucl. Med.; (United States) · OSTI ID:6699410

Use of O-15 water and C-11 butanol to measure cerebral blood flow (CBF) and water permeability with positron emission tomography (PET)
Conference · Wed May 01 00:00:00 EDT 1985 · J. Nucl. Med.; (United States) · OSTI ID:6851740