Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Optimizing the Operation of a Vertical Johann Spectrometer Using a High Energy Fluorescer X-ray Source

Conference · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.3492421· OSTI ID:1136444
This paper describes the operation and testing for a Vertical Johann Spectrometer (VJS) operating in the 13 keV range. The spectrometer is designed to use thin curved mica crystals or thick germanium crystals. The VJS must have a resolution E/ΔE=3000 or better to measure Doppler broadening of highly ionized krypton and operate at a small X-ray angle in order to be used as a diagnostic in a laser plasma target chamber. The VJS was aligned, tested, and optimized using a fluorescer type high energy X-ray (HEX) source located at National Security Technologies, LLC (NSTec), in Livermore, California. The HEX uses a 160 kV X-ray tube to excite fluorescence from various targets. Both rubidium and bismuth fluorescers were used for this effort. This presentation describes the NSTec HEX system and the methods used to optimize and characterize the VJS performance.
Research Organization:
Nevada Test Site/National Security Technologies, LLC (United States)
Sponsoring Organization:
USDOE National Nuclear Security Administration (NNSA)
DOE Contract Number:
AC52-06NA25946
OSTI ID:
1136444
Report Number(s):
DOE/NV/25946--990
Conference Information:
Journal Name: Review of Scientific Instruments Journal Issue: 10 Journal Volume: 81
Country of Publication:
United States
Language:
English

Similar Records

Optimizing the operation of a high resolution vertical Johann spectrometer using a high energy fluorescer x-ray source
Journal Article · Fri Oct 15 00:00:00 EDT 2010 · Review of Scientific Instruments · OSTI ID:22055903

Measuring Curved Crystal Performance for a High Resolution, Imaging X-ray Spectrometer
Journal Article · Mon Jun 07 00:00:00 EDT 2010 · X-Ray Optics and Instrumentation · OSTI ID:993431

High-resolution compact Johann crystal spectrometer with the Livermore electron beam ion trap
Journal Article · Fri Oct 01 00:00:00 EDT 2004 · Review of Scientific Instruments · OSTI ID:20636731

Related Subjects