Title: Source calibrations and SDC calorimeter requirements

Technical Report ·
DOI:https://doi.org/10.2172/6969889· OSTI ID:6969889

Several studies of the problem of calibration of the SDC calorimeter exist. In this note the attempt is made to give a connected account of the requirements on the source calibration from the point of view of the desired, and acceptable, constant term induced in the EM resolution. It is assumed that a local'' calibration resulting from exposing each tower to a beam of electrons is not feasible. It is further assumed that an in situ'' calibration is either not yet performed, or is unavailable due to tracking alignment problems or high luminosity operation rendering tracking inoperative. Therefore, the assumptions used are rather conservative. In this scenario, each scintillator plate of each tower is exposed to a moving radioactive source. That reading is used to mask'' an optical cookie'' in a grey code chosen so as to make the response uniform. The source is assumed to be the sole calibration of the tower. Therefore, the phrase global'' calibration of towers by movable radioactive sources is adopted.