skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Automated detection and reporting of Volatile Organic Compounds (VOCs) in complex environments

Technical Report ·
DOI:https://doi.org/10.2172/469115· OSTI ID:469115

This paper describes results from efforts to develop VOC sensing systems based on two complementary techniques. The first technique used a gated channeltron detector for resonant laser-induced multiphoton photoionization detection of trace organic vapors in a supersonic molecular beam. The channeltron was gated using a relatively simple circuit to generate a negative gate pulse with a width of 400 ns (FWHM), a 50 ns turn-on (rise) time, a 1.5 {mu}s turn-off (decay) time, a pulse amplitude of {minus}1000 Volts, and a DC offset adjustable from zero to {minus}1500 Volts. The gated channeltron allows rejection of spurious responses to UV laser light scattered directly into the channeltron and time-delayed ionization signals induced by photoionization of residual gas in the vacuum chamber. Detection limits in the part-per-trillion range have been demonstrated with the gated detector. The second technique used arrays of surface acoustic wave (SAW) devices coated with various chemically selective materials (e.g., polymers, self assembled monolayers) to provide unique response patterns to various chemical analytes. This work focused on polymers, formed by spin casting from solution or by plasma polymerization, as well as on self assembled monolayers. Response from coated SAWs to various concentrations of water, volatile organics, and organophosphonates (chemical warfare agent simulants) were used to provide calibration data. A novel visual empirical region of influence (VIERI) pattern recognition technique was used to evaluate the ability to use these response patterns to correctly identify chemical species. This investigation shows how the VERI technique can be used to determine the best set of coatings for an array, to predict the performance of the array even if sensor responses change due to aging of the coating materials, and to identify unknown analytes based on previous calibration data.

Research Organization:
Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
Sponsoring Organization:
USDOE Assistant Secretary for Environment, Safety, and Health, Washington, DC (United States)
DOE Contract Number:
AC04-94AL85000
OSTI ID:
469115
Report Number(s):
SAND-97-0509; ON: DE97005908; TRN: 97:002969
Resource Relation:
Other Information: PBD: Mar 1997
Country of Publication:
United States
Language:
English