Title: SU-E-T-533: Evaluation of Dose Calculation Accuracy for Small Elongated Targets On the Edge Linac

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888867· OSTI ID:22369751
; ; ; ; ; ; ; ; ; ;  [1]
  1. Henry Ford Health System, Detroit, MI (United States)

Purpose: To evaluate output factors and dose calculation accuracy on a novel SRS linear accelerator, the Edge (Varian), for treatments of small, elongated targets using flattening filter free (FFF) beam. Methods: Total scatter/output factors (OF’s) for 24 elongated, small, high definition multi-leaf collimator (HDMLC)-defined fields were measured on the Edge machine using 6X FFF beam. 3 detectors were used in water tank: CC01 ion chamber (active volume 10cc), stereotactic photon diode (SFD) (active diameter 0.6mm, active thickness 0.06mm), Edge detector (active volume 0.0019cc). The 24 MLC apertures have widths ranging from 5 to 20mm and length/width ratio from 0.25 to 5. Readings were cross calibrated with CC04 at field size 3×3 cm. A beam model was developed using commissioning measurements for treatment planning in Eclipse (AAA, version 11). One representative patient case (IMRT, target volume 0.2cc, 4×4×14mm) was calculated using AAA 11 and delivered on the Edge. Results: Due to volume averaging effects, CC01 readings were 11.2±0.9% lower than SFD readings for 5mm field sizes. The Edge diode showed a uniform over-response of 2.6±0.7% compared to SFD. Calculation using AAA v11 showed the best agreement with SFD measurements (2.4±1.7% lower than SFD). The largest difference between AAA v11 and SFD occurs at 5mm field sizes. For the patient plan, dose delivered on Edge was measured to be 2.2% higher than AAA v11 calculation. Conclusion: Cross-calibrated SFD output measurements presented the best agreement with commissioned AAA v11 beam model. Field sizes smaller than 1cm posed challenges to both the detectors and the calculation algorithm. For the representative patient with small elongated target, AAA v11 and measurements agreed within ~2% on the Edge linac. Although encouraging, a more comprehensive study is required to validate the overall algorithmic accuracy.

OSTI ID:
22369751
Journal Information:
Medical Physics, Vol. 41, Issue 6; Other Information: (c) 2014 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English

Similar Records

SU-G-TeP2-08: Evaluation of Plastic Scintillator Detector for Small Field Stereotactic Patient-Specific Quality Assurance
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22369751

True beam commissioning experience at Nordland Hospital Trust, Norway
Journal Article · Fri Mar 25 00:00:00 EDT 2016 · AIP Conference Proceedings · OSTI ID:22369751

SU-E-T-370: Measurement of Conical Cone Output Factors for the Varian Edge Linear Accelerator
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22369751