SU-E-T-370: Measurement of Conical Cone Output Factors for the Varian Edge Linear Accelerator

Journal Article · · Medical Physics
DOI:https://doi.org/10.1118/1.4888703· OSTI ID:22355915
; ; ;  [1];  [2];  [3];  [4]
  1. Henry Ford Health System, Detroit, MI (United States)
  2. Wayne State University School of Medicine, Detroit, MI (United States)
  3. Henry Ford Hospital System, Detroit, MI (United States)
  4. Henry Ford Health System, West Bloomfield, MI (United States)

Purpose: To quantify the impact of detector type, SSD/depth, and intermediate reference on conical cone output factor (OF) measurements for the Varian Edge linac. Methods: OF's for 4, 5, 7.5, 10, 12.5, 15, and 17.5 mm diameter cones relative to 10cmx10cm field were measured for the 6X FFF and 10X FFF energies, with jaws set to 5cmx5cm. Measurements were performed with an Edge diode (0.8mmx0.8mmx0.03mm WxLxT), stereotatic diode SFD, photon diode, CC01 and pinpoint chambers (2mm diameter for both). 95cm SSD/5cm depth were used in a water tank. For the measurement with diodes, OF's were cross-referred to CC13 ion chamber measurements with 3cmx3cm field, as recommended, to help mitigate the energy variation in diode response with field size. Results were compared to the representative data from Varian measured with Edge detector. With SFD, OF's at 98.5cm SSD/1.5cm depth and 90cm SSD/10cm depth were also measured. Results: OF's measured with the Edge detector matched within 1.3% (max diff) with the representative data from Varian. For the SFD, OF's matched within 1.3% for the 4, 5 and 17.5 mm cones and within 3.7% for the other cones. OF's with photon diode were within 1.3% except for the 4 and 5 mm cones where they were 8.1% and 3.7%, respectively. OF's for the CC01 and pinpoint chamber deviated up to 36% and 44%, respectively for the 4 mm cone. OF's after intermediate reference with 3cmx3cm field changed by 3.7% for SFD, 0.8% for photon diode, and 0.6% for Edge detector. OF's at 98.5cm SSD/1.5cm depth were 10.8% higher than that at 95cm SSD/5cm depth, and OF's at 90cm SSD/1.5cm depth were 7.5% lower. Conclusion: OF's measured with the Edge detector appear to be reliable. CC01 and pinpoint chambers do not appear suitable for measuring the small cone OF's. SSD/depth affects OF measurements significantly.

OSTI ID:
22355915
Journal Information:
Medical Physics, Journal Name: Medical Physics Journal Issue: 6 Vol. 41; ISSN 0094-2405; ISSN MPHYA6
Country of Publication:
United States
Language:
English

Similar Records

SU-E-T-533: Evaluation of Dose Calculation Accuracy for Small Elongated Targets On the Edge Linac
Journal Article · Sun Jun 01 00:00:00 EDT 2014 · Medical Physics · OSTI ID:22369751

SU-F-T-572: Small Fields Output Factors Measurement and Correction Factor Determination for Several Active and Passive Detectors
Journal Article · Wed Jun 15 00:00:00 EDT 2016 · Medical Physics · OSTI ID:22649147

SU-E-T-246: Detector Selection for Small Field Measurements, a Comparative Study
Journal Article · Mon Jun 15 00:00:00 EDT 2015 · Medical Physics · OSTI ID:22548304