In vivo measurements for high dose rate brachytherapy with optically stimulated luminescent dosimeters
- Department of Radiation Oncology, West Michigan Cancer Center, 200 North Park Street, Kalamazoo, Michigan 49007 (United States)
Purpose: To show the feasibility of clinical implementation of OSLDs for high dose-rate (HDR) in vivo dosimetry for gynecological and breast patients. To discuss how the OSLDs were characterized for an Ir-192 source, taking into account low gamma energy and high dose gradients. To describe differences caused by the dose calculation formalism of treatment planning systems.Methods: OSLD irradiations were made using the GammaMedplus iX Ir-192 HDR, Varian Medical Systems, Milpitas, CA. BrachyVision versions 8.9 and 10.0, Varian Medical Systems, Milpitas, CA, were used for calculations. Version 8.9 used the TG-43 algorithm and version 10.0 used the Acuros algorithm. The OSLDs (InLight Nanodots) were characterized for Ir-192. Various phantoms were created to assess calculated and measured doses and the angular dependence and self-absorption of the Nanodots. Following successful phantom measurements, patient measurements for gynecological patients and breast cancer patients were made and compared to calculated doses.Results: The OSLD sensitivity to Ir-192 compared to 6 MV is between 1.10 and 1.25, is unique to each detector, and changes with accumulated dose. The measured doses were compared to those predicted by the treatment planning system and found to be in agreement for the gynecological patients to within measurement uncertainty. The range of differences between the measured and Acuros calculated doses was -10%-14%. For the breast patients, there was a discrepancy of -4.4% to +6.5% between the measured and calculated doses at the skin surface when the Acuros algorithm was used. These differences were within experimental uncertainty due to (random) error in the location of the detector with respect to the treatment catheter.Conclusions: OSLDs can be successfully used for HDR in vivo dosimetry. However, for the measurements to be meaningful one must account for the angular dependence, volume-averaging, and the greater sensitivity to Ir-192 gamma rays than to 6 MV x-rays if 6 MV x-rays were used for OSLD calibration. The limitations of the treatment planning algorithm must be understood, especially for surface dose measurements. Use of in vivo dosimetry for HDR brachytherapy treatments is feasible and has the potential to detect and prevent gross errors. In vivo HDR brachytherapy should be included as part of the QA for a HDR brachytherapy program.
- OSTI ID:
- 22121624
- Journal Information:
- Medical Physics, Vol. 40, Issue 7; Other Information: (c) 2013 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
- Country of Publication:
- United States
- Language:
- English
Similar Records
Development and implementation of a remote audit tool for high dose rate (HDR) Ir-192 brachytherapy using optically stimulated luminescence dosimetry
SU-F-BRA-12: End-User Oriented Tools and Procedures for Testing Brachytherapy TPSs Employing MBDCAs
Related Subjects
62 RADIOLOGY AND NUCLEAR MEDICINE
ALGORITHMS
BRACHYTHERAPY
CALIBRATION
DOSE RATES
DOSIMETRY
GAMMA RADIATION
IN VIVO
IRIDIUM 192
LUMINESCENCE
MAMMARY GLANDS
NEOPLASMS
PHANTOMS
PLANNING
QUANTUM DOTS
RADIATION DOSES
SELF-ABSORPTION
SKIN
THERMOLUMINESCENT DOSEMETERS
X RADIATION