Purpose: To develop user-oriented tools for commissioning and dosimetry testing of {sup 192}Ir brachytherapy treatment planning systems (TPSs) employing model based dose calculation algorithms (MBDCAs). Methods: A software tool (BrachyGuide) has been developed for the automatic generation of MCNP6 input files from any CT based plan exported in DICOM RT format from Elekta and Varian TPSs. BrachyGuide also facilitates the evaluation of imported Monte Carlo (MC) and TPS dose distributions in terms of % dose differences and gamma index (CT overlaid colormaps or relative frequency plots) as well as DVHs and related indices. For users not equipped to perform MC, a set of computational models was prepared in DICOM format, accompanied by treatment plans and corresponding MCNP6 generated reference data. BrachyGuide can then be used to compare institutional and reference data as per TG186. The model set includes a water sphere with the MBDCA WG {sup 192}Ir source placed centrically and in two eccentric positions, a water sphere with cubic bone and lung inhomogeneities and a five source dwells plan, and a patient equivalent model with an Accelerated Partial Breast Irradiation (APBI) plan. Results: The tools developed were used for the dosimetry testing of the Acuros and ACE MBDCAs implemented in BrachyVision v.13 and Oncentra Brachy v.4.5, respectively. Findings were consistent with previous results in the literature. Besides points close to the source dwells, Acuros was found to agree within type A uncertainties with the reference MC results. Differences greater than MC type A uncertainty were observed for ACE at distances >5cm from the source dwells and in bone. Conclusion: The tools developed are efficient for brachytherapy MBDCA planning commissioning and testing. Since they are appropriate for distribution over the web, they will be put at the AAPM WG MBDCA’s disposal. Research co-financed by the ESF and Greek funds. NSRF operational Program: Education and Lifelong Learning Investing in Knowledge Society-Aristeia. Varian Medical Systems and Nucletron, an Elekta company provided access to TPSs for research purposes. Miss Peppa was supported by IKY-fellowships of excellence for postgraduate studies in Greece,Siemens Program.

OSTI ID:
22555218
Journal Information:
Medical Physics, Vol. 42, Issue 6; Other Information: (c) 2015 American Association of Physicists in Medicine; Country of input: International Atomic Energy Agency (IAEA); ISSN 0094-2405
Country of Publication:
United States
Language:
English