DOE PAGES title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: High energy X-ray pinhole imaging at the Z facility

Abstract

A new high photon energy (hν > 15 keV) time-integrated pinhole camera (TIPC) has been developed as a diagnostic instrument at the Z facility. This camera employs five pinholes in a linear array for recording five images at once onto an image plate detector. Each pinhole may be independently filtered to yield five different spectral responses. The pinhole array is fabricated from a 1-cm thick tungsten block and is available with either straight pinholes or conical pinholes. Each pinhole within the array block is 250 μm in diameter. The five pinholes are splayed with respect to each other such that they point to the same location in space, and hence present the same view of the radiation source at the Z facility. The fielding distance from the radiation source is 66 cm and the nominal image magnification is 0.374. Initial experimental results from TIPC are shown to illustrate the performance of the camera.

Authors:
ORCiD logo [1]; ORCiD logo [1];  [1];  [1];  [1]; ORCiD logo [1]
  1. Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
Publication Date:
Research Org.:
Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)
Sponsoring Org.:
USDOE National Nuclear Security Administration (NNSA)
OSTI Identifier:
1259535
Alternate Identifier(s):
OSTI ID: 1420628
Report Number(s):
SAND-2015-10135J
Journal ID: ISSN 0034-6748; RSINAK; 608469
Grant/Contract Number:  
AC04-94AL85000; 165733; 173104
Resource Type:
Accepted Manuscript
Journal Name:
Review of Scientific Instruments
Additional Journal Information:
Journal Volume: 87; Journal Issue: 6; Journal ID: ISSN 0034-6748
Publisher:
American Institute of Physics (AIP)
Country of Publication:
United States
Language:
English
Subject:
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY

Citation Formats

McPherson, L. Armon, Ampleford, David J., Coverdale, Christine A., Argo, Jeffrey W., Owen, Albert C., and Jaramillo, Deanna M. High energy X-ray pinhole imaging at the Z facility. United States: N. p., 2016. Web. doi:10.1063/1.4953004.
McPherson, L. Armon, Ampleford, David J., Coverdale, Christine A., Argo, Jeffrey W., Owen, Albert C., & Jaramillo, Deanna M. High energy X-ray pinhole imaging at the Z facility. United States. https://doi.org/10.1063/1.4953004
McPherson, L. Armon, Ampleford, David J., Coverdale, Christine A., Argo, Jeffrey W., Owen, Albert C., and Jaramillo, Deanna M. Mon . "High energy X-ray pinhole imaging at the Z facility". United States. https://doi.org/10.1063/1.4953004. https://www.osti.gov/servlets/purl/1259535.
@article{osti_1259535,
title = {High energy X-ray pinhole imaging at the Z facility},
author = {McPherson, L. Armon and Ampleford, David J. and Coverdale, Christine A. and Argo, Jeffrey W. and Owen, Albert C. and Jaramillo, Deanna M.},
abstractNote = {A new high photon energy (hν > 15 keV) time-integrated pinhole camera (TIPC) has been developed as a diagnostic instrument at the Z facility. This camera employs five pinholes in a linear array for recording five images at once onto an image plate detector. Each pinhole may be independently filtered to yield five different spectral responses. The pinhole array is fabricated from a 1-cm thick tungsten block and is available with either straight pinholes or conical pinholes. Each pinhole within the array block is 250 μm in diameter. The five pinholes are splayed with respect to each other such that they point to the same location in space, and hence present the same view of the radiation source at the Z facility. The fielding distance from the radiation source is 66 cm and the nominal image magnification is 0.374. Initial experimental results from TIPC are shown to illustrate the performance of the camera.},
doi = {10.1063/1.4953004},
journal = {Review of Scientific Instruments},
number = 6,
volume = 87,
place = {United States},
year = {Mon Jun 06 00:00:00 EDT 2016},
month = {Mon Jun 06 00:00:00 EDT 2016}
}

Journal Article:
Free Publicly Available Full Text
Publisher's Version of Record

Citation Metrics:
Cited by: 13 works
Citation information provided by
Web of Science

Save / Share:

Works referenced in this record:

Status of the Z pulsed power driver
conference, June 2011


Enhancement of X-Ray Power from a Z Pinch Using Nested-Wire Arrays
journal, November 1998


Experimental Demonstration of Fusion-Relevant Conditions in Magnetized Liner Inertial Fusion
journal, October 2014


A Renewed Capability for Gas Puff Science on Sandia's Z Machine
journal, May 2014

  • Jones, Brent; Jennings, Christopher A.; Lamppa, Derek C.
  • IEEE Transactions on Plasma Science, Vol. 42, Issue 5
  • DOI: 10.1109/TPS.2013.2287180

Compact, rugged in-chamber transmission spectrometers (7–28 keV) for the Sandia Z facility
journal, June 2011

  • Sinars, D. B.; Wenger, D. F.; Pikuz, S. A.
  • Review of Scientific Instruments, Vol. 82, Issue 6
  • DOI: 10.1063/1.3600610

Multilayer mirror monochromatic self-emission x-ray imaging on the Z accelerator
journal, October 2006

  • Jones, B.; Deeney, C.; Coverdale, C. A.
  • Review of Scientific Instruments, Vol. 77, Issue 10
  • DOI: 10.1063/1.2220071

Pinned, optically aligned diagnostic dock for use on the Z facility
journal, October 2012

  • Gomez, M. R.; Rochau, G. A.; Bailey, J. E.
  • Review of Scientific Instruments, Vol. 83, Issue 10
  • DOI: 10.1063/1.4732848

Contrasting physics in wire array z pinch sources of 1-20 keV emission on the Z facility
journal, May 2014

  • Ampleford, D. J.; Jones, B.; Jennings, C. A.
  • Physics of Plasmas, Vol. 21, Issue 5
  • DOI: 10.1063/1.4876621

Imaging of high-energy x-ray emission from cryogenic thermonuclear fuel implosions on the NIF
journal, October 2012

  • Ma, T.; Izumi, N.; Tommasini, R.
  • Review of Scientific Instruments, Vol. 83, Issue 10
  • DOI: 10.1063/1.4733313

Implications of high-energy photons and electrons on target preheat at the Sandia “Z” facility
journal, January 1999

  • Derzon, M.; Nash, T.; Chandler, G.
  • Review of Scientific Instruments, Vol. 70, Issue 1
  • DOI: 10.1063/1.1149445

Opacity and gradients in aluminum wire array z-pinch implosions on the Z pulsed power facility
journal, March 2014

  • Ampleford, D. J.; Hansen, S. B.; Jennings, C. A.
  • Physics of Plasmas, Vol. 21, Issue 3
  • DOI: 10.1063/1.4865224

K-shell emission trends from 60 to 130 cm/ μ s stainless steel implosions
journal, October 2013

  • Ampleford, D. J.; Jennings, C. A.; Jones, B.
  • Physics of Plasmas, Vol. 20, Issue 10
  • DOI: 10.1063/1.4823711

A New Method of Spectroscopy for Faint X-Radiations
journal, January 1928


Analysis and implementation of a space resolving spherical crystal spectrometer for x-ray Thomson scattering experiments
journal, April 2015

  • Harding, E. C.; Ao, T.; Bailey, J. E.
  • Review of Scientific Instruments, Vol. 86, Issue 4
  • DOI: 10.1063/1.4918619

High-energy x-ray response of photographic films: models and measurement
journal, January 1986

  • Henke, B. L.; Stone, G. F.; Fujiwara, F. G.
  • Journal of the Optical Society of America B, Vol. 3, Issue 11
  • DOI: 10.1364/JOSAB.3.001540

Evaluation of the sensitivity and fading characteristics of an image plate system for x-ray diagnostics
journal, November 2008

  • Meadowcroft, A. L.; Bentley, C. D.; Stott, E. N.
  • Review of Scientific Instruments, Vol. 79, Issue 11
  • DOI: 10.1063/1.3013123

Signatures of hot electrons and fluorescence in Mo Kα emission on Z
journal, March 2014

  • Hansen, S. B.; Ampleford, D. J.; Cuneo, M. E.
  • Physics of Plasmas, Vol. 21, Issue 3
  • DOI: 10.1063/1.4866161

Works referencing / citing this record:

Coarse spectral characterization of warm x-rays at the Z facility using a filtered thermoluminescent dosimeter array
journal, April 2017

  • Harper-Slaboszewicz, V. J.; Ulmen, B. A.; Parzyck, C. T.
  • Review of Scientific Instruments, Vol. 88, Issue 4
  • DOI: 10.1063/1.4979626

A Wolter imager on the Z machine to diagnose warm x-ray sources
journal, October 2018

  • Fein, J. R.; Ampleford, D. J.; Vogel, J. K.
  • Review of Scientific Instruments, Vol. 89, Issue 10
  • DOI: 10.1063/1.5038347

Design and raytrace simulations of a multilayer-coated Wolter x-ray optic for the Z machine at Sandia National Laboratories
journal, October 2018

  • Vogel, J. K.; Pivovaroff, M. J.; Kozioziemski, B.
  • Review of Scientific Instruments, Vol. 89, Issue 10
  • DOI: 10.1063/1.5038811