Optimization and application of cooled avalanche photodiodes for spectroscopic fluctuation measurements with ultra-fast charge exchange recombination spectroscopy
Abstract
The Ultra Fast Charge Exchange Recombination Spectroscopy (UF-CHERS) diagnostic is a highly specialized spectroscopic instrument with 2 spatial channels consisting of 8 spectral channels each and a resolution of ~0.25 nm deployed at DIII-D to measure turbulent ion temperature fluctuations. Charge exchange emissions are obtained between 528-530 nm with 1 μs time resolution to study plasma instabilities. A primary challenge of extracting fluctuation measurements from raw UF-CHERS signals is photon and electronic noise. In order to reduce dark current, the Avalanche Photodiode (APD) detectors are thermoelectrically cooled. State-of-the-art components are used for the signal amplifiers and conditioners to minimize electronic noise. Due to the low incident photon power (≤ 1 nW), APDs with a gain of up to 300 are used to optimize the signal to noise ratio. Maximizing the APDs’ gain while minimizing the excess noise factor (ENF) is essential since the total noise of the diagnostic sets a floor for the minimum level of detectable broadband fluctuations. The APDs’ gain should be high enough that photon noise dominates electronic noise, but not excessive so that the ENF overwhelms plasma fluctuations. A new generation of cooled APDs and optimized preamplifiers exhibits significantly enhanced signal-to-noise compared to a previous generation.more »
- Authors:
-
- Univ. of Wisconsin, Madison, WI (United States)
- Publication Date:
- Research Org.:
- Univ. of Wisconsin System, Madison, WI (United States)
- Sponsoring Org.:
- USDOE
- Contributing Org.:
- Department of Engineering Physics, University of Wisconsin – Madison, Madison, WI 53706, USA
- OSTI Identifier:
- 1371908
- Grant/Contract Number:
- FG02-89ER53296
- Resource Type:
- Accepted Manuscript
- Journal Name:
- Review of Scientific Instruments
- Additional Journal Information:
- Journal Volume: 87; Journal Issue: 11; Related Information: D.D. Truong, R.J. Fonck, and G.R. McKee, "Optimization and application of cooled avalance photodiodes for spectroscopic fluctuation measurements with ultra-fast charge exchange recombination spectroscopy," Rev. Sci. Instr. 87, 11E551 (2016).; Journal ID: ISSN 0034-6748
- Publisher:
- American Institute of Physics (AIP)
- Country of Publication:
- United States
- Language:
- English
- Subject:
- 70 PLASMA PHYSICS AND FUSION TECHNOLOGY; 46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
Citation Formats
Truong, D. D., Fonck, R. J., and McKee, G. R. Optimization and application of cooled avalanche photodiodes for spectroscopic fluctuation measurements with ultra-fast charge exchange recombination spectroscopy. United States: N. p., 2016.
Web. doi:10.1063/1.4963147.
Truong, D. D., Fonck, R. J., & McKee, G. R. Optimization and application of cooled avalanche photodiodes for spectroscopic fluctuation measurements with ultra-fast charge exchange recombination spectroscopy. United States. https://doi.org/10.1063/1.4963147
Truong, D. D., Fonck, R. J., and McKee, G. R. Fri .
"Optimization and application of cooled avalanche photodiodes for spectroscopic fluctuation measurements with ultra-fast charge exchange recombination spectroscopy". United States. https://doi.org/10.1063/1.4963147. https://www.osti.gov/servlets/purl/1371908.
@article{osti_1371908,
title = {Optimization and application of cooled avalanche photodiodes for spectroscopic fluctuation measurements with ultra-fast charge exchange recombination spectroscopy},
author = {Truong, D. D. and Fonck, R. J. and McKee, G. R.},
abstractNote = {The Ultra Fast Charge Exchange Recombination Spectroscopy (UF-CHERS) diagnostic is a highly specialized spectroscopic instrument with 2 spatial channels consisting of 8 spectral channels each and a resolution of ~0.25 nm deployed at DIII-D to measure turbulent ion temperature fluctuations. Charge exchange emissions are obtained between 528-530 nm with 1 μs time resolution to study plasma instabilities. A primary challenge of extracting fluctuation measurements from raw UF-CHERS signals is photon and electronic noise. In order to reduce dark current, the Avalanche Photodiode (APD) detectors are thermoelectrically cooled. State-of-the-art components are used for the signal amplifiers and conditioners to minimize electronic noise. Due to the low incident photon power (≤ 1 nW), APDs with a gain of up to 300 are used to optimize the signal to noise ratio. Maximizing the APDs’ gain while minimizing the excess noise factor (ENF) is essential since the total noise of the diagnostic sets a floor for the minimum level of detectable broadband fluctuations. The APDs’ gain should be high enough that photon noise dominates electronic noise, but not excessive so that the ENF overwhelms plasma fluctuations. A new generation of cooled APDs and optimized preamplifiers exhibits significantly enhanced signal-to-noise compared to a previous generation. Experiments at DIII-D have allowed for characterization and optimization of the ENF vs. gain. Here, a gain of ~100 at 1700 V is found to be near optimal for most plasma conditions. Ion temperature and toroidal velocity fluctuations due to the Edge Harmonic Oscillation (EHO) in Quiescent H-mode (QH) plasmas are presented to demonstrate UF-CHERS’ capabilities.},
doi = {10.1063/1.4963147},
journal = {Review of Scientific Instruments},
number = 11,
volume = 87,
place = {United States},
year = {Fri Sep 23 00:00:00 EDT 2016},
month = {Fri Sep 23 00:00:00 EDT 2016}
}
Web of Science
Works referenced in this record:
Low‐noise photodiode detector for optical fluctuation diagnostics
journal, October 1992
- Fonck, R. J.; Ashley, R.; Durst, R.
- Review of Scientific Instruments, Vol. 63, Issue 10
Low-noise, high-speed detector development for optical turbulence fluctuation measurements for NSTX
journal, October 2010
- Schoenbeck, N. L.; Ellington, S. D.; Fonck, R. J.
- Review of Scientific Instruments, Vol. 81, Issue 10
Ultrafast ion temperature and toroidal velocity fluctuation spectroscopy diagnostic design
journal, October 2008
- McKee, G. R.; Schlossberg, D. J.; Shafer, M. W.
- Review of Scientific Instruments, Vol. 79, Issue 10
A review of experimental drift turbulence studies
journal, October 2009
- Tynan, G. R.; Fujisawa, A.; McKee, G.
- Plasma Physics and Controlled Fusion, Vol. 51, Issue 11
Ultrafast spectroscopy diagnostic to measure localized ion temperature and toroidal velocity fluctuations
journal, October 2010
- Uzun-Kaymak, I. U.; Fonck, R. J.; McKee, G. R.
- Review of Scientific Instruments, Vol. 81, Issue 10
Enhanced sensitivity beam emission spectroscopy system for nonlinear turbulence measurements
journal, October 2004
- Gupta, Deepak K.; Fonck, Raymond J.; McKee, George R.
- Review of Scientific Instruments, Vol. 75, Issue 10
Drift waves and transport
journal, April 1999
- Horton, W.
- Reviews of Modern Physics, Vol. 71, Issue 3
Chapter 2: Plasma confinement and transport
journal, June 2007
- Physics), E. J. Doyle (Chair Transport; Modelling), W. A. Houlberg (Chair Confinement Da; Edge), Y. Kamada (Chair Pedestal and
- Nuclear Fusion, Vol. 47, Issue 6
Ultra-fast charge exchange spectroscopy for turbulent ion temperature fluctuation measurements on the DIII-D tokamak (invited)
journal, October 2012
- Uzun-Kaymak, I. U.; Fonck, R. J.; McKee, G. R.
- Review of Scientific Instruments, Vol. 83, Issue 10