Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Diffraction-Enhanced Imaging of Musculoskeletal Tissues Using a Conventional X-Ray Tube

Journal Article · · Academic Radiology
DEI based on a conventional x-ray tube allows the visualization of skeletal and soft tissues simultaneously. Although more in-depth testing and optimization of the DEI setup must be carried out, these data demonstrate a proof of principle for further development of the technology for future clinical imaging. In conventional projection radiography, cartilage and other soft tissues do not produce enough radiographic contrast to be distinguishable from each other. Diffraction-enhanced imaging (DEI) uses a monochromatic x-ray beam and a silicon crystal analyzer to produce images in which attenuation contrast is greatly enhanced and x-ray refraction at tissue boundaries can be detected. The aim of this study was to test the efficacy of conventional x-ray tube-based DEI for the detection of soft tissues in experimental samples.
Research Organization:
Brookhaven National Laboratory (BNL) National Synchrotron Light Source
Sponsoring Organization:
Doe - Office Of Science
DOE Contract Number:
AC02-98CH10886
OSTI ID:
980150
Report Number(s):
BNL--93068-2010-JA
Journal Information:
Academic Radiology, Journal Name: Academic Radiology Journal Issue: 8 Vol. 16
Country of Publication:
United States
Language:
English

Similar Records

Design and Implementation of a Compact Low-Dose Diffraction Enhanced Medical Imaging System
Journal Article · Wed Dec 31 23:00:00 EST 2008 · Academic Radiology · OSTI ID:980135

Multiple-Image Radiography for Human Soft Tissue
Journal Article · Sat Dec 31 23:00:00 EST 2005 · J. Anatomy · OSTI ID:914395

Radiologist Evaluation of an X-ray Tube Based Diffraction Enhanced Imaging Prototype Using Full Thickness Breast Specimens.