Improved DESI-MS Performance using Edge Sampling and aRotational Sample Stage
- ORNL
The position of the surface to be analyzed relative to the sampling orifice or capillary into the mass spectrometer has been known to dramatically affect the observed signal levels in desorption electrospray ionization mass spectrometry (DESIMS). In analyses of sample spots on planar surfaces, DESI-MS signal intensities as much as five times greater were routinely observed when the bottom of the sampling capillary was appropriately positioned beneath the surface plane ( edge sampling") compared to when the capillary just touched the surface. To take advantage of the optimum "edge sampling" geometry and to maximize the number of samples that could be analyzed in this configuration, a rotational sample stage was integrated into a typical DESI-MS setup. The rapid quantitative determination of caffeine in two diet sport drinks (Diet Turbo Tea, Speed Stack Grape) spiked with an isotopically labeled internal standard demonstrated the utility of this approach.
- Research Organization:
- Oak Ridge National Laboratory (ORNL)
- Sponsoring Organization:
- ORNL work for others; SC USDOE - Office of Science (SC)
- DOE Contract Number:
- AC05-00OR22725
- OSTI ID:
- 959440
- Journal Information:
- Rapid Communications in Mass Spectrometry, Journal Name: Rapid Communications in Mass Spectrometry Journal Issue: 23 Vol. 22; ISSN RCMSEF; ISSN 0951-4198
- Country of Publication:
- United States
- Language:
- English
Similar Records
Quantitative Thin-Layer Chromatography/Mass Spectrometry Analysis of Caffeine Using a Surface Sampling Probe Electrospray Ionization Tandem Mass Spectrometry System
Field Testing of the COTS DESI/DART-MS/MS System Against High Explosives Residues