skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Enhanced Edot X System

Abstract

The capability of EdotX would be improved by increasing the instrument's dynamic range to expand the range of signals measurable by the device. The heart of the EdotX system is a two channel analog to digital (A/D) conversion card with two analog to digital converters tightly coupled to a field programmable gate array (FPGA). Signals are continually digitized by the A/D and processed by the FPGA. When the FPGA detects a trigger condition, the data are transferred using direct memory access to the host computer memory, which in turn saves the data to disk. The original design utilizes a single A/D channel for data acquisition. The task was to increase the dynamic range of the system by enabling the second A/D channel, allowing the simultaneous acquisition of the original A/D channel and the additional second A/D channel. The input signal would be fed to the original channel and to the second channel with the second channel input being attenuated. When a signal hits the maximum range of the first channel, the data from the second channel, which is attenuated, are then used. Because both A/D's acquire data simultaneously, they may be used as a single expanded dynamic range channel.

Authors:
;
Publication Date:
Research Org.:
UNLV Research Foundation, 8311 W. Sunset Road, Suite 200, Las Vegas, NV 89113
Sponsoring Org.:
USDOE National Nuclear Security Administration (NA)
OSTI Identifier:
957048
Report Number(s):
DOE/NA/27513-F
TRN: US201018%%333
DOE Contract Number:
FG52-06NA27513
Resource Type:
Technical Report
Country of Publication:
United States
Language:
English
Subject:
99 GENERAL AND MISCELLANEOUS; ANALOG-TO-DIGITAL CONVERTERS; COMPUTERS; DATA ACQUISITION; DESIGN; EdotX; EDOTX; data acquisition equipment; two channel analog to digital conversion car; field programmable gate array

Citation Formats

Smith, Scott, and Harry Bostick. Enhanced Edot X System. United States: N. p., 2009. Web. doi:10.2172/957048.
Smith, Scott, & Harry Bostick. Enhanced Edot X System. United States. doi:10.2172/957048.
Smith, Scott, and Harry Bostick. Mon . "Enhanced Edot X System". United States. doi:10.2172/957048. https://www.osti.gov/servlets/purl/957048.
@article{osti_957048,
title = {Enhanced Edot X System},
author = {Smith, Scott and Harry Bostick},
abstractNote = {The capability of EdotX would be improved by increasing the instrument's dynamic range to expand the range of signals measurable by the device. The heart of the EdotX system is a two channel analog to digital (A/D) conversion card with two analog to digital converters tightly coupled to a field programmable gate array (FPGA). Signals are continually digitized by the A/D and processed by the FPGA. When the FPGA detects a trigger condition, the data are transferred using direct memory access to the host computer memory, which in turn saves the data to disk. The original design utilizes a single A/D channel for data acquisition. The task was to increase the dynamic range of the system by enabling the second A/D channel, allowing the simultaneous acquisition of the original A/D channel and the additional second A/D channel. The input signal would be fed to the original channel and to the second channel with the second channel input being attenuated. When a signal hits the maximum range of the first channel, the data from the second channel, which is attenuated, are then used. Because both A/D's acquire data simultaneously, they may be used as a single expanded dynamic range channel.},
doi = {10.2172/957048},
journal = {},
number = ,
volume = ,
place = {United States},
year = {Mon Jun 15 00:00:00 EDT 2009},
month = {Mon Jun 15 00:00:00 EDT 2009}
}

Technical Report:

Save / Share:
  • An autonomous gas chromatograph system was designed and built to support the Thermal Enhanced Vapor Extraction System (TEVES) demonstration. TEVES is a remediation demonstration that seeks to enhance an existing technology (vacuum extraction) by adding a new technology (soil heating). A pilot scale unit was set up at one of the organic waste disposal pits at the Sandia National Laboratories Chemical Waste Landfill (CWL) in Tech Area 3. The responsibility for engineering a major part of the process instrumentation for TEVES belonged to the Manufacturing Control Subsystems Department. The primary mission of the one-of-a-kind hardware/software system is to perform on-sitemore » gas sampling and analysis to quantify a variety of volatile organic compounds (VOCs) from various sources during TEVES operations. The secondary mission is to monitor a variety of TEVES process physical parameters such as extraction manifold temperature, pressure, humidity, and flow rate, and various subsurface pressures. The system began operation in September 1994 and was still in use on follow-on projects when this report was published.« less
  • Substantial unexploited opportunity exists for the US, and the world, in Enhanced Geothermal Systems (EGS). As a result of US DOE investment, new drilling technology, new power generation equipment and cycles enable meaningful power production, in a compact and modular fashion; at lower and lower top side EGS working fluid temperatures and in a broader range of geologies and geographies. This cost analysis effort supports the expansion of Enhanced Geothermal Systems (EGS), furthering DOE strategic themes of energy security and sub goal of energy diversity; reducing the Nation's dependence on foreign oil while improving the environment.
  • This document describes the Enhanced Automated System that is designed for use during the program planning phase of the Department of Energy (DOE) FY83-87 Planning, Programming and Budgeting System (PPBS) cycle. The Enhanced Automated Syst
  • This report describes an evaluation study for the application of stereo-X-radiography as a tool for three dimensional visualization and quantitative assessment of mechanical damage in graphite-epoxy structural materials. Comparison of image quality produced using X-ray opaque liquid penetrant materials was made for stereo-X-radiographs of panels that contained mechanical damage. A new X-ray opaque penetrant material was formulated and used for development of a high resolution stereo-X-radiographic technique. The technique was used to document damage accumulation in a series of pre-damaged panels during incremental tension-tension fatigue loading. The high resolution technique was shown to be useful in assessment of accumulated damagemore » and in quantitative location of the damage within a panel. Features of the technique include use of a new X-ray opaque penetrant formulation, use of a unique penetrameter, 'soft radiation' and a fine grained X-ray film.« less
  • High quality YBa2Cu3O7-x films were grown in-situ on LaAlO3 (100) by a novel single liquid source plasma-enhanced metalorganic chemical vapor deposition process. The metalorganic complexes M(thd)n, (thd = 2,2,6,6-tetramethyl-3,5-heptanedionate; M = Y, Ba, Cu) were dissolved in an organic solution and injected into a vaporizer immediately upstream of the reactor inlet The single liquid source technique dramatically simplifies current CVD processing and can significantly improve the process reproducibility. X-ray diffraction. measurements indicated that single phase, highly c-axis oriented YBa2Cu3O7-x was formed in-situ at a substrate temperature 680 degC. The as-deposited films exhibited a mirror-like surface, had transition temperature Tc =more » 89 K, Delta Tc < 1K, and Jc(77K) = 106 A/cm2. Plasma enhanced metalorganic chemical vapor deposition, YBCO, superconductors.« less