A Bremsstrahlung Spectrometer using k-edge and Differential Filters with Image plate dosimeters
A Bremsstrahlung spectrometer using k-edge and differential filtering has been used with Image Plate dosimeters to measure the x-ray fluence from short-pulse laser/target interactions. An electron spectrometer in front of the Bremsstrahlung spectrometer deflects electrons from the x-ray line of sight and simultaneously measures the electron spectrum. The response functions were modeled with the Monte Carlo code Integrated Tiger Series 3.0 and the dosimeters calibrated with radioactive sources. Electron distributions with slope temperatures in the MeV range are inferred from the Bremsstrahlung spectra.
- Research Organization:
- Lawrence Livermore National Laboratory (LLNL), Livermore, CA
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- W-7405-ENG-48
- OSTI ID:
- 938491
- Report Number(s):
- LLNL-CONF-403592
- Country of Publication:
- United States
- Language:
- English
Similar Records
A Bremsstrahlung spectrometer using k-edge and differential filters with image plate dosimeters
Experimental verification of non-equilibrated bremsstrahlung dosimetry predictions for 0. 75 MeV electrons
Accuracy evaluation of a Compton X-ray spectrometer with bremsstrahlung X-rays generated by a 6 MeV electron bunch
Journal Article
·
Wed Oct 15 04:00:00 UTC 2008
· Review of Scientific Instruments
·
OSTI ID:21266470
Experimental verification of non-equilibrated bremsstrahlung dosimetry predictions for 0. 75 MeV electrons
Conference
·
Wed Jan 01 04:00:00 UTC 1986
·
OSTI ID:5842291
Accuracy evaluation of a Compton X-ray spectrometer with bremsstrahlung X-rays generated by a 6 MeV electron bunch
Journal Article
·
Sat Nov 15 04:00:00 UTC 2014
· Review of Scientific Instruments
·
OSTI ID:22303468