Accuracy evaluation of a Compton X-ray spectrometer with bremsstrahlung X-rays generated by a 6 MeV electron bunch
Journal Article
·
· Review of Scientific Instruments
- Institute of Laser Engineering, Osaka University, 2-6 Yamada-oka, Suita, Osaka 565-0871 (Japan)
- Toyota Technical Development Corporation, 1-21 Imae, Hanamoto-cho, Toyota, Aichi 470-0334 (Japan)
- National Institute for Fusion Science, 322-6 Oroshicho, Toki, Gifu 509-5292 (Japan)
- The Institute of Science and Industrial Research, Osaka University, 2-6 Yamada-oka, Suita, Osaka (Japan)
A Compton-scattering-based X-ray spectrometer is developed to obtain the energy distribution of fast electrons produced by intense laser and matter interactions. Bremsstrahlung X-rays generated by fast electrons in a material are used to measure fast electrons’ energy distribution in matter. In the Compton X-ray spectrometer, X-rays are converted into recoil electrons by Compton scattering in a converter made from fused silica glass, and a magnet-based electron energy analyzer is used to measure the energy distribution of the electrons that recoil in the direction of the incident X-rays. The spectrum of the incident X-rays is reconstructed from the energy distribution of the recoil electrons. The accuracy of this spectrometer is evaluated using a quasi-monoenergetic 6 MeV electron bunch that emanates from a linear accelerator. An electron bunch is injected into a 1.5 mm thick tungsten plate to produce bremsstrahlung X-rays. The spectrum of these bremsstrahlung X-rays is obtained in the range from 1 to 9 MeV. The energy of the electrons in the bunch is estimated using a Monte Carlo simulation of particle-matter interactions. The result shows that the spectrometer's energy accuracy is ±0.5 MeV for 6.0 MeV electrons.
- OSTI ID:
- 22303468
- Journal Information:
- Review of Scientific Instruments, Journal Name: Review of Scientific Instruments Journal Issue: 11 Vol. 85; ISSN 0034-6748; ISSN RSINAK
- Country of Publication:
- United States
- Language:
- English
Similar Records
Photonuclear reaction based high-energy x-ray spectrometer to cover from 2 MeV to 20 MeV
Development of Compton X-ray spectrometer for high energy resolution single-shot high-flux hard X-ray spectroscopy
X-ray spectrometer for observation of nonlinear compton scattering.
Journal Article
·
Fri Nov 14 23:00:00 EST 2014
· Review of Scientific Instruments
·
OSTI ID:22303694
Development of Compton X-ray spectrometer for high energy resolution single-shot high-flux hard X-ray spectroscopy
Journal Article
·
Fri Apr 15 00:00:00 EDT 2016
· Review of Scientific Instruments
·
OSTI ID:22597125
X-ray spectrometer for observation of nonlinear compton scattering.
Conference
·
Wed Dec 31 23:00:00 EST 2003
·
OSTI ID:925080