Subambient pressure ionization with nanoelectrospray source and interface for improved sensitivity in mass spectrometry

Journal Article · · Analytical Chemistry, Vol. 80(5):1800-1805
DOI:https://doi.org/10.1021/ac702354b· OSTI ID:926105
An electrospray ionization mass spectrometry (ESI-MS) source and interface has been designed that enables efficient ion production and transmission in a 30 Torr pressure environment using solvents compatible with typical reverse-phase liquid chromatography (RPLC) separations. In this design, the electrospray emitter is located inside the mass spectrometer in the same region as an electrodynamic ion funnel. This avoids the use of a conductance limit ion inlet, as required by a conventional atmospheric pressure ESI source, and allows more efficient ion transmission to the mass analyzer. Performance of the low pressure ESI source and interface was evaluated by electrospraying standard solutions at 300 nL/min, and comparing results with those obtained from a standard atmospheric pressure ESI source that used a heated capillary inlet. The importance of desolvation was also investigated by electrospraying at different flow rates, which showed that the ion funnel provided an effective desolvation region to aid the creation of gas phase analyte ions. This initial study demonstrated a ~ 5-fold improvement in sensitivity when the low pressure ESI source was used compared to a standard atmospheric pressure ESI source.
Research Organization:
Pacific Northwest National Laboratory (PNNL), Richland, WA (US)
Sponsoring Organization:
USDOE
DOE Contract Number:
AC05-76RL01830
OSTI ID:
926105
Report Number(s):
PNNL-SA-57906; KP1601010
Journal Information:
Analytical Chemistry, Vol. 80(5):1800-1805, Journal Name: Analytical Chemistry, Vol. 80(5):1800-1805 Journal Issue: 5 Vol. 80
Country of Publication:
United States
Language:
English

Similar Records

A subambient pressure ionization with nanoelectrospray (SPIN) source and interface for improved sensitivity in mass spectrometry
Journal Article · Sat Mar 15 00:00:00 EDT 2008 · Analytical Chemistry, 80(5):1800-1805 · OSTI ID:1013315

Improving liquid chromatography-mass spectrometry sensitivity using a subambient pressure ionization with nanoelectrospray (SPIN) interface
Journal Article · Fri Apr 22 00:00:00 EDT 2011 · Journal of the American Society for Mass Spectrometry, 22(8):1318-1325 · OSTI ID:1020628

Improving mass spectrometer sensitivity using a high-pressure electrodynamic ion funnel interface
Journal Article · Fri Sep 15 00:00:00 EDT 2006 · Journal of the American Society for Mass Spectrometry · OSTI ID:22774019

Related Subjects