skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Calibration of an isotopically enriched carbon-13 layer pressure sensor to 156 GPa in a diamond anvil cell

Journal Article · · J. Appl. Phys.
DOI:https://doi.org/10.1063/1.2184436· OSTI ID:914177

An isotopically enriched {sup 13}C homoepitaxial diamond layer of 6{+-}1 {mu}m thickness was grown on top of a brilliant cut diamond anvil by a microwave plasma chemical vapor deposition process for application as a pressure sensor. This isotopically enriched diamond tip was then used in conjunction with a natural isotopic abundance diamond anvil to generate high pressure on the sample. We provide a calibration for the {sup 13}C Raman mode of this extremely thin epitaxial layer to 156 GPa using ruby fluorescence and the equation of state of copper as secondary pressure standards. The nonlinear calibration of the {sup 13}C Raman mode pressure sensor is compared with similar calibrations of {sup 12}C Raman edge and a good agreement is obtained. The Raman signal from the {sup 13}C epitaxial layer remained a distinct singlet to 156 GPa, and pressure calibration is independent of sample mechanical strength or the diamond anvil geometry. The use of even thinner layer would allow calibration further into ultrahigh pressure regime where the use of other optical sensors has proven to be difficult.

Research Organization:
Brookhaven National Lab. (BNL), Upton, NY (United States). National Synchrotron Light Source
Sponsoring Organization:
Doe - Office Of Science
DOE Contract Number:
DE-AC02-98CH10886
OSTI ID:
914177
Report Number(s):
BNL-78745-2007-JA; JAPIAU; TRN: US200804%%577
Journal Information:
J. Appl. Phys., Vol. 99; ISSN 0021-8979
Country of Publication:
United States
Language:
English