Absolute Calibration of Kodak Biomax-MS Film Response to X Rays in the 1.5- to 8-keV Energy Range
Conference
·
· Review of Scientific Instruments
- Laboratory for Laser Energetics, University of Rochester, Rochester, NY
The absolute response of Kodak Biomax-MS film to x rays in the range from 1.5- to 8-keV has been measured using a laboratory e-beam generated x-ray source. The measurements were taken at specific line energies by using Bragg diffraction to produce monochromatic beams of x rays. Multiple exposures were taken on Biomax MS film up to levels exceeding optical densities of 2 as measured by a microdensitometer. The absolute beam intensity for each exposure was measured with a Si(Li) detector. Additional response measurements were taken with Kodak direct exposure film (DEF) so as to compare the results of this technique to previously published calibrations.
- Research Organization:
- Laboratory for Laser Energetics, University of Rochester
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- FC52-92SF19460
- OSTI ID:
- 892219
- Report Number(s):
- DOE/SF/19460-700; 1662; 2006-20
- Conference Information:
- Journal Name: Review of Scientific Instruments Journal Volume: 77
- Country of Publication:
- United States
- Language:
- English
Similar Records
Absolute calibration of Kodak Biomax-MS film response to x rays in the 1.5- to 8-keV energy range
Response Model for Kodak Biomax-MS Film to X Rays
Response model for Kodak Biomax-MS film to x rays
Journal Article
·
Sun Oct 15 00:00:00 EDT 2006
· Review of Scientific Instruments
·
OSTI ID:20861338
Response Model for Kodak Biomax-MS Film to X Rays
Journal Article
·
Tue Jan 23 23:00:00 EST 2007
· Review of Scientific Instruments
·
OSTI ID:897882
Response model for Kodak Biomax-MS film to x rays
Journal Article
·
Sun Oct 15 00:00:00 EDT 2006
· Review of Scientific Instruments
·
OSTI ID:20861361