skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: Response model for Kodak Biomax-MS film to x rays

Journal Article · · Review of Scientific Instruments
DOI:https://doi.org/10.1063/1.2220046· OSTI ID:20861361
; ; ; ; ; ; ; ; ;  [1]
  1. Laboratory for Laser Energetics, University of Rochester, Rochester, New York 14623 (United States)

X-ray-sensitive film is used for a variety of imaging and spectroscopic diagnostics for high-temperature plasmas. Replacement film must be found as older films are phased out of production. Biomax-MS is a 'T-grain' class of film that is proposed as a replacement for Kodak DEF and a model of its response to x rays is presented. Data from dimensional measurements of the film, x-ray transmission measurements, scanning electron microscopy micrograph images, and x-ray calibration are used to develop this sensitivity model of Biomax-MS film as a function of x-ray energy and angle of incidence. Relative response data provide a check of the applicability of this model to determine the x-ray flux from spectrum data. This detailed film characterization starts with simple mathematical models and extends them to T-grain-type film.

OSTI ID:
20861361
Journal Information:
Review of Scientific Instruments, Vol. 77, Issue 10; Other Information: DOI: 10.1063/1.2220046; (c) 2006 American Institute of Physics; Country of input: International Atomic Energy Agency (IAEA); ISSN 0034-6748
Country of Publication:
United States
Language:
English