Improving the Accuracy of Using Pyranometers to Measure the Clear Sky Global Solar Irradiance
Pyranometer users have customarily applied one responsivity value when calculating the global solar irradiance. Usually, the responsivity value is reported by either the manufacturer or a calibration facility. Many pyranometer calibrations, made both at NREL and elsewhere, have shown that the responsivity of a pyranometer changes with the change in solar zenith and azimuth angles. Depending on how well the pyranometer sensor is radiometrically leveled, these changes can exceed +/-5% of the reported responsivity, which means that errors in the calculated global solar irradiance can exceed +/-5% from the nominal values. This paper describes a method to decrease the errors resulting from the change of the solar zenith angle under clear sky conditions. Two responsivity functions, morning and afternoon, were used instead of one responsivity value. The two functions have been chosen because of asymmetry of the morning and afternoon cosine responses demonstrated by some pyranometers.
- Research Organization:
- National Renewable Energy Laboratory (NREL), Golden, CO.
- Sponsoring Organization:
- USDOE
- DOE Contract Number:
- AC36-99GO10337
- OSTI ID:
- 882188
- Report Number(s):
- NREL/TP-560-24833
- Country of Publication:
- United States
- Language:
- English
Similar Records
Uncertainty Estimate for the Outdoor Calibration of Solar Pyranometers: A Metrologist Perspective
The Impact of Indoor and Outdoor Radiometer Calibration on Solar Measurements
Calculating the Diffuse Responsivity of Solar Pyranometers
Journal Article
·
Sun Nov 30 23:00:00 EST 2008
· Measure
·
OSTI ID:981980
The Impact of Indoor and Outdoor Radiometer Calibration on Solar Measurements
Conference
·
Wed Feb 01 23:00:00 EST 2017
·
OSTI ID:1343074
Calculating the Diffuse Responsivity of Solar Pyranometers
Technical Report
·
Mon Aug 16 00:00:00 EDT 1999
·
OSTI ID:9811