Calibration of a PEM detector with depth of interactionmeasurement

Journal Article · · IEEE Transactions on Nuclear Science

We present an in situ calibration technique for the LBNL positron emission mammography (PEM) detector module that is capable of measuring depth of interaction (DOI). The detector module consists of 64LSO crystals coupled on one end to a single photomultiplier tube (PMT) and on the opposite end to a 64 pixel array of silicon photodiodes (PD). The PMT provides an accurate timing pulse, the PDs identify the crystal of interaction, the sum provides a total energy signal and the /splGamma/=PD/(PD+PMT) ratio determines the depth of interaction. We calibrate using the /sup 176/Lu natural background radiation of the LSO crystals. We determine the relative gain (K) of the PMT and PD by minimizing the asymmetry of the /spl Gamma/ distribution. We determine the depth dependence from the width of the /spl Gamma/ distribution with optimal K. The performance of calibrated detector modules is evaluated by averaging results from 12 modules. The energy resolution is a function of depth ranging from 24 percent FWHM at the PD end to 51 percent FWHM at the PMT end, and the DOI resolution ranges from 6 mm FWHM at the PD end to 11 mm FWHM at the PMT end.

Research Organization:
Ernest Orlando Lawrence Berkeley NationalLaboratory, Berkeley, CA (US)
Sponsoring Organization:
USDOE Director, Office of Science. Office of Biological andEnvironmental Research. Medical Sciences Division; National Institutes ofHealth
DOE Contract Number:
AC02-05CH11231
OSTI ID:
881601
Report Number(s):
LBNL--52708; BnR: 400412000
Journal Information:
IEEE Transactions on Nuclear Science, Journal Name: IEEE Transactions on Nuclear Science Journal Issue: 3 Vol. 51; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English