A LSO scintillator array for a PET detector module with depth of interaction measurement

Conference ·
OSTI ID:786960

We present construction methods and performance results for a production scintillator array of 64 optically isolated, 3 mm x 3 mm x 30 mm sized LSO crystals. This scintillator array has been developed for a PET detector module consisting of the 8x8 LSO array coupled on one end to a single photomultiplier tube (PMT) and on the opposite end to a 64 pixel array of silicon photodiodes (PD). The PMT provides an accurate timing pulse and initial energy discrimination, the PD identifies the crystal of interaction, the sum provides a total energy signal, and the PD/(PD+PMT) ratio determines the depth of interaction (DOI). Unlike the previous LSO array prototypes, we now glue Lumirror reflector material directly onto 4 sides of each crystal to obtain an easily manufactured, mechanically rugged array with our desired depth dependence. With 511 keV excitation, we obtain a total energy signal of 3600 electrons, pulse-height resolution of 25% fwhm, and 6-15 mm fwhm DOI resolution.

Research Organization:
Ernest Orlando Lawrence Berkeley National Laboratory, Berkeley, CA (US)
Sponsoring Organization:
USDOE Director, Office of Science. Office of Biological and Environmental Research. Medical Sciences Division; National Institutes of Health (US)
DOE Contract Number:
AC03-76SF00098
OSTI ID:
786960
Report Number(s):
LBNL--47067
Country of Publication:
United States
Language:
English