Performance of a PSPMT based detector for scintimammography
In breast scintigraphy, compact detectors with high intrinsic spatial resolution and small inactive peripheries can provide improvements in extrinsic spatial resolution, efficiency and contrast for small lesions relative to larger conventional cameras. We are developing a pixelated small field-of-view gamma camera for scintimammography. Extensive measurements of the imaging properties of a prototype system have been made, including spatial resolution, sensitivity, uniformity of response, geometric linearity and energy resolution. An anthropomorphic torso phantom providing a realistic breast exit gamma spectrum has been used in a qualitative study of lesion detectability. A new type of breast imaging system that combines scintimammography and digital mammography in a single upright unit has also been developed. The system provides automatic co-registration between the scintigram and the digital mammogram, obtained with the breast in a single configuration.
- Research Organization:
- Thomas Jefferson National Accelerator Facility, Newport News, VA (US)
- Sponsoring Organization:
- USDOE Office of Energy Research (ER) (US)
- DOE Contract Number:
- AC05-84ER40150
- OSTI ID:
- 804874
- Report Number(s):
- JLAB-PHY-00-56; DOE/ER/40150-2391
- Journal Information:
- Phys. Med. Biol., Journal Name: Phys. Med. Biol. Journal Issue: 3 Vol. 45
- Country of Publication:
- United States
- Language:
- English
Similar Records
A Combined Scintimammography/Stereotactic Core Biopsy Digital X-ray System
X-ray stereotactic lesion localization in conjunction with dedicated scintimammography