High speed gel permeation chromatography calibration procedures. Period covered: January--March 1976

Technical Report ·
DOI:https://doi.org/10.2172/7269227· OSTI ID:7269227
A High Speed Gel Permeation Chromatograph (HSGPC) calibration curve was generated by use of approximately 160 calibration runs employing polystyrene and hydrocarbon standards which were fitted by a fifth degree polynomial. Three substances, the peroxide of tetrahydrofuran, o-dichlorobenzene, and HMX were evaluated as internal standards to correct for small fluctuations in flow rate from run-to-run, and were found to be satisfactory. The utility of a composite calibration mixture was also established.
Research Organization:
Mason and Hanger-Silas Mason Co., Inc., Amarillo, Tex. (USA)
OSTI ID:
7269227
Report Number(s):
MHSMP-76-17W
Country of Publication:
United States
Language:
English