Performance testing of extremity dosimeters, Study 2
- Pacific Northwest Lab., Richland, WA (USA)
The Health Physics Society Standards Committee (HPSSC) Working Group on Performance Testing of Extremity Dosimeters has issued a draft of a proposed standard for extremity dosimeters. The draft standard proposes methods to be used for testing dosimetry systems that determine occupational radiation dose to the extremities and the performance criterion used to determine compliance with the standard. Pacific Northwest Laboratory (PNL) has conducted two separate evaluations of the performance of extremity dosimeter processors to determine the appropriateness of the draft standard, as well as to obtain information regarding the performance of extremity dosimeters. Based on the information obtained during the facility visits and the results obtained from the performance testing, it was recommended that changes be made to ensure that the draft standard is appropriate for extremity dosimeters. The changes include: subdividing the mixture category and the beta particle category; eliminating the neutron category until appropriate flux-to-dose equivalent conversion factors are derived; and changing the tolerance level for the performance criterion to provide consistency with the performance criterion for whole body dosimeters, and to avoid making the draft standard overly difficult for processors of extremity dosimeters to pass. 20 refs., 10 figs., 6 tabs.
- Research Organization:
- Nuclear Regulatory Commission, Washington, DC (USA). Div. of Regulatory Applications; Pacific Northwest Lab., Richland, WA (USA)
- Sponsoring Organization:
- NRC
- DOE Contract Number:
- AC06-76RL01830
- OSTI ID:
- 7199063
- Report Number(s):
- NUREG/CR-5540; PNL--7276; ON: TI90009928
- Country of Publication:
- United States
- Language:
- English
Similar Records
Performance testing of extremity dosimeters
Performance testing of extremity dosimeters -- pilot test
Performance testing of extremity dosimeters -- pilot test
Technical Report
·
Mon Jun 01 00:00:00 EDT 1987
·
OSTI ID:6352251
Performance testing of extremity dosimeters -- pilot test
Technical Report
·
Thu Jul 01 00:00:00 EDT 1993
·
OSTI ID:10173079
Performance testing of extremity dosimeters -- pilot test
Technical Report
·
Thu Jul 01 00:00:00 EDT 1993
·
OSTI ID:6282716
Related Subjects
440102* -- Radiation Instrumentation-- Radiation Dosemeters
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
BARYONS
BETA PARTICLES
CALIBRATION
CHARGED PARTICLES
COMPILED DATA
COMPLIANCE
DATA
DOSEMETERS
DOSES
ELEMENTARY PARTICLES
EQUATIONS
ERRORS
EXTERNAL IRRADIATION
FERMIONS
HADRONS
INFORMATION
IRRADIATION
MASSLESS PARTICLES
MEASURING INSTRUMENTS
NEUTRONS
NUCLEONS
NUMERICAL DATA
OCCUPATIONAL EXPOSURE
PERFORMANCE TESTING
PERSONNEL
PHOTONS
RADIATION DOSES
STANDARDS
TESTING
WHOLE-BODY IRRADIATION
46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY
BARYONS
BETA PARTICLES
CALIBRATION
CHARGED PARTICLES
COMPILED DATA
COMPLIANCE
DATA
DOSEMETERS
DOSES
ELEMENTARY PARTICLES
EQUATIONS
ERRORS
EXTERNAL IRRADIATION
FERMIONS
HADRONS
INFORMATION
IRRADIATION
MASSLESS PARTICLES
MEASURING INSTRUMENTS
NEUTRONS
NUCLEONS
NUMERICAL DATA
OCCUPATIONAL EXPOSURE
PERFORMANCE TESTING
PERSONNEL
PHOTONS
RADIATION DOSES
STANDARDS
TESTING
WHOLE-BODY IRRADIATION