Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Design and performance of a new positron tomograph

Conference · · IEEE Trans. Nucl. Sci.; (United States)
OSTI ID:7029012
High spatial resolution in a positron tomograph with multislice imaging capability can be achieved by some form of position encoding to avoid the complexity resulting from individual photomultipliers (PMT) coupled to individual small crystals. The authors recently completed a multislice positron tomograph using six large area NaI (T1) crystals coupled to 50-mm square PMTs, which results in a large number of resolution elements with relatively low complexity compared to other approaches. The design and performance of this instrument, which is based on a previously constructed single-slice version, is described in this paper, emphasizing those aspects which differ from other designs.
Research Organization:
Univ. of Pennsylvania, Dept. of Radiology, 3400 Spruce Street, Philadelphia, PA (US)
OSTI ID:
7029012
Report Number(s):
CONF-871006-
Conference Information:
Journal Name: IEEE Trans. Nucl. Sci.; (United States) Journal Volume: 35:1
Country of Publication:
United States
Language:
English

Similar Records

Event localization in a continuous scintillation detector using digital processing
Conference · Fri Jan 31 23:00:00 EST 1986 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:5886541

Multislice positron emission computed tomograph (PETT IV) yielding transverse and longitudinal images
Conference · Tue Aug 01 00:00:00 EDT 1978 · Radiology; (United States) · OSTI ID:6952384

Performance characteristics of the McMaster positron emission tomograph
Conference · Tue Jan 31 23:00:00 EST 1984 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:6850532