We designed, built, and tested a positron emission tomograph (PETT IV) capable of providing seven slices of the human body simultaneously. PETT IV utilizes a moving hexagonal array of 48 scintillation detectors placed around the subject. Each detector consists of a cylindrical activated sodium iodide crystal optically coupled to two photomultiplier tubes. The multislice capability is achieved by comparing the light outputs of the two photomultiplier tubes in each detector. The images are displayed either as transverse or as longitudinal tomographic sections. This system provides high sensitivity and resolution, and permits rapid and accurate three-dimensional imaging of the head and body.

Research Organization:
Univ. of Washington, St. Louis
OSTI ID:
6952384
Journal Information:
Radiology; (United States), Vol. 128:2; Conference: 63. scientific assembly and annual meeting of the Radiological Society of North America, Chicago, IL, USA, 26 Nov 1977
Country of Publication:
United States
Language:
English

Similar Records

Engineering developments on the UBC-Triumf modified PETT VI positron emission tomograph
Journal Article · Tue Feb 01 00:00:00 EST 1983 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:6952384

Performance characteristics of newly developed whole-body multislice positron computed tomograph
Conference · Sun Jan 01 00:00:00 EST 1984 · J. Nucl. Med.; (United States) · OSTI ID:6952384

Event localization in a continuous scintillation detector using digital processing
Conference · Sat Feb 01 00:00:00 EST 1986 · IEEE Trans. Nucl. Sci.; (United States) · OSTI ID:6952384