A ruggedized ZnS(Ag)/epoxy alpha scintillation detector

Conference ·
OSTI ID:6943879

An alpha scintillation survey instrument has been developed which is more rugged and efficient than conventional alpha scintillation detectors that use aluminized Mylar{dagger} radiation entrance windows. This new detector consists of a mixture of ZnS(Ag) phosphor and optically transparent epoxy. The scintillator mixture is poured into a preformed mold to provide a thin layer of phosphor after the particles settle to the clear epoxy surface. After partial curing, an optically transparent light pipe is coupled to the ZnS(Ag)/epoxy film by using an additional thin epoxy layer, forming a monolithic scintillator assembly. 2 refs., 3 figs., 2 tabs.