Hip prostheses during pelvic irradiation: effects and corrections
Journal Article
·
· Int. J. Radiat. Oncol., Biol. Phys.; (United States)
Treatment of pelvic malignancies frequently includes the use of lateral, arc, or rotational fields. The presence of hip prostheses in these treatment fields will perturb the dose distribution. Correction factors for metal-based alloys used in artificial hips have not previously been reported. Prostheses constructed from frequently used alloys were obtained and measurements were made of the transmission of 4MV and 10MV photons. These measured data were compared with computed correction factors. The computer uses the ratio of tissue-maximum ratios (TMR's) method of heterogeneity correction. The computer was provided with both the physical density and the relative electron density of each prosthesis for comparison purposes, since electron densities for hip prostheses are not widely known. Correction factors determined from electron densities demonstrated better agreement with measured data. The ratio of TMR's correction algorithm does not consider the contribution of scattered radiation in the dose computations. Consequently, a small adjustment to the relative electron density of the prosthetic hip was required at lower X ray beam energies. Agreement was satisfactory for higher energy X rays, and thus no adjustment was necessary. Relative electron densities and adjusted electron densities for alloys used in artificial hips are provided for computer-aided treatment planning. Recommendations for incorporating the hip prosthesis into the treatment planning process are also provided.
- Research Organization:
- Presbyterian/Saint Luke's Medical Center, Denver, CO (USA)
- OSTI ID:
- 6908493
- Journal Information:
- Int. J. Radiat. Oncol., Biol. Phys.; (United States), Journal Name: Int. J. Radiat. Oncol., Biol. Phys.; (United States) Vol. 14:6; ISSN IOBPD
- Country of Publication:
- United States
- Language:
- English
Similar Records
Prostatic thermoluminescent dosimeter analysis in a patient treated with 18 MV X rays through a prosthetic hip
Radiotherapy dose calculations in the presence of hip prostheses
Experimental and Monte Carlo evaluation of Eclipse treatment planning system for effects on dose distribution of the hip prostheses
Journal Article
·
Thu Jan 14 23:00:00 EST 1993
· International Journal of Radiation Oncology, Biology and Physics; (United States)
·
OSTI ID:6613860
Radiotherapy dose calculations in the presence of hip prostheses
Journal Article
·
Mon Jun 30 00:00:00 EDT 2003
· Medical Dosimetry
·
OSTI ID:20655446
Experimental and Monte Carlo evaluation of Eclipse treatment planning system for effects on dose distribution of the hip prostheses
Journal Article
·
Tue Oct 01 00:00:00 EDT 2013
· Medical Dosimetry
·
OSTI ID:22262832
Related Subjects
550603* -- Medicine-- External Radiation in Therapy-- (1980-)
62 RADIOLOGY AND NUCLEAR MEDICINE
BODY
BODY AREAS
DISEASES
DOSES
ELECTROMAGNETIC RADIATION
ELECTRONS
ELEMENTARY PARTICLES
ELEMENTS
FERMIONS
IONIZING RADIATIONS
IRRADIATION PROCEDURES
LEPTONS
MEDICAL SUPPLIES
MEDICINE
METALS
NEOPLASMS
NUCLEAR MEDICINE
PELVIS
PLANNING
PROSTHESES
RADIATION DOSES
RADIATIONS
RADIOLOGY
RADIOTHERAPY
RECOMMENDATIONS
THERAPY
TITANIUM
TRANSITION ELEMENTS
X RADIATION
62 RADIOLOGY AND NUCLEAR MEDICINE
BODY
BODY AREAS
DISEASES
DOSES
ELECTROMAGNETIC RADIATION
ELECTRONS
ELEMENTARY PARTICLES
ELEMENTS
FERMIONS
IONIZING RADIATIONS
IRRADIATION PROCEDURES
LEPTONS
MEDICAL SUPPLIES
MEDICINE
METALS
NEOPLASMS
NUCLEAR MEDICINE
PELVIS
PLANNING
PROSTHESES
RADIATION DOSES
RADIATIONS
RADIOLOGY
RADIOTHERAPY
RECOMMENDATIONS
THERAPY
TITANIUM
TRANSITION ELEMENTS
X RADIATION