Local threshold for segmented attenuation correction of PET imaging of the thorax

Conference · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
OSTI ID:6906776
;  [1];  [2];  [3]
  1. Mallinckrodt Inst. of Radiology, St. Louis, MO (United States)
  2. Mallinckrodt Inst. of Radiology, St. Louis, MO (United States) CTI-PET System, Inc., Knoxville, TN (United States)
  3. CTI-PET System, Inc., Knoxville, TN (United States)

A local threshold for segmented attenuation correction technique has been developed for positron emission tomography using short post-injection transmission scans. The technique implements an optimal threshold method on localized histograms to get pseudo-anatomic segmentation on transmission images. Theoretical values of attenuation coefficients are assigned to corresponding anatomic regions. Emission images are reconstructed using attenuation correction factors computed by forward-projecting segmented transmission images. Phantoms and clinical cardiac images are studied using this technique. The technique corrects emission images with accuracy similar to the standard, pre-injection method, reduces noise in the corrected emission images, and offers the potential for increased patient throughput by enabling shorter, post-injection data acquisition.

OSTI ID:
6906776
Report Number(s):
CONF-931051--
Journal Information:
IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States), Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) Vol. 41:4PT1; ISSN 0018-9499; ISSN IETNAE
Country of Publication:
United States
Language:
English

Similar Records

Local threshold segmented attenuation correction with emission data subtraction for PET thorax imaging
Journal Article · Sun May 01 00:00:00 EDT 1994 · Journal of Nuclear Medicine · OSTI ID:197924

PET attenuation correction using reprojection of measured transmission data
Conference · Sat Dec 31 23:00:00 EST 1983 · J. Nucl. Med.; (United States) · OSTI ID:6779512

Object-specific attenuation correction of SPECT with correlated dual-energy X-ray CT
Journal Article · Sun Aug 01 00:00:00 EDT 1993 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:5695718