skip to main content
OSTI.GOV title logo U.S. Department of Energy
Office of Scientific and Technical Information

Title: The HgI sub 2 energy dispersive x-ray array detectors and minaturized processing electronics project

Conference · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA)
OSTI ID:6872257
; ; ;  [1];  [2]; ;  [3]
  1. University of Southern California, Marina Del Rey, CA (USA). Inst. of Physics
  2. X-Ray Instrumentation Associates, Menlo Park, CA (US)
  3. Stanford Univ., CA (USA). Stanford Synchrotron Radiation Lab.

This paper describes recent progress in the development of HgI{sub 2} energy dispersive x-ray detector arrays for synchrotron radiation research and their associated miniaturized processing electronics. Deploying a 5 element HgI{sub 2} array detector under realistic operating conditions at SSRL, an energy resolution of 252 eV FWHM at 5.9 keV (Mn-K{alpha}) was obtained. The authors also report energy resolution and throughput measurements versus input count rate. The results from the HgI{sub 2} system are then compared to those obtained under identical conditions from a commercial 13 element Ge detector array.

OSTI ID:
6872257
Report Number(s):
CONF-900143-; CODEN: IETNA; TRN: 90-031754
Journal Information:
IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA), Vol. 37:2; Conference: Institute for Electronic and Electrical Engineers (IEEE) nuclear science symposium, San Francisco, CA (USA), 15-19 Jan 1990; ISSN 0018-9499
Country of Publication:
United States
Language:
English