Title: 20-element HgI[sub 2] energy dispersive x-ray array detector system

Conference · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
OSTI ID:6731254
; ; ; ;  [1]; ;  [2];  [3]
  1. Xsirius, Inc., Marina del Ray, CA (United States)
  2. Stanford Univ., CA (United States). Stanford Synchrotron Radiation Lab.
  3. EG and G Energy Measurements, Inc., Goleta, CA (United States)

This paper describes recent progress in the development of HgI[sub 2] energy dispersive x-ray arrays and associated miniaturized processing electronics for synchrotron radiation research applications. The experimental results with a 20-element array detector were obtained under realistic synchrotron beam conditions at SSRL. An energy resolution of 250 eV (FWHM) at 5.9 keV (Mn-K[sub alpha]) was achieved. Energy resolution and throughput measurements versus input count rate and energy of incoming radiation have been measured. Extended X-ray Absorption Fine Structure (EXAFS) spectra were taken form diluted samples simulating proteins with nickel.

OSTI ID:
6731254
Report Number(s):
CONF-911106-; CODEN: IETNAE; TRN: 93-006856
Journal Information:
IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States), Vol. 39:5; Conference: 1991 Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference, Santa Fe, NM (United States), 2-9 Nov 1991; ISSN 0018-9499
Country of Publication:
United States
Language:
English