Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Improving the performance of the SP-3000 PET detector modules

Conference · · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States)
OSTI ID:6855289
; ;  [1]
  1. Washington Univ., Seattle, WA (United States)
This paper reports on a major concern in the use of BaF[sub 2] detectors in a positron emission tomograph which is the collection of the UV light from the fast decay component of the scintillator. A series of experiments with the standard SP-3000 commercial detector module confirmed that the coupling compound formulation was reasonable, but that the reflector material used to wrap the crystals was not optimal. Rebuilding the detector modules with PTFE Teflon increased the light output from the detector modules by factors of three. Initial tests with phantoms using two rings of rebuilt detectors in the University of Washington BaF[sub 2] PET system confirmed improvements in resolution and contrast in comparison to the original SP-3000 modules.
OSTI ID:
6855289
Report Number(s):
CONF-911106--
Conference Information:
Journal Name: IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) Journal Volume: 39:4
Country of Publication:
United States
Language:
English

Similar Records

A new preprocessor for the SP-3000 series of time-of-flight positron emission tomographs
Journal Article · Sun Aug 01 00:00:00 EDT 1993 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (United States) · OSTI ID:6011795

High resolution block detectors for PET
Conference · Sat Mar 31 23:00:00 EST 1990 · IEEE Transactions on Nuclear Science (Institute of Electrical and Electronics Engineers); (USA) · OSTI ID:6606784

Studies with a prototype high resolution PET scanner based on LSO-APD modules
Journal Article · Mon Jun 01 00:00:00 EDT 1998 · IEEE Transactions on Nuclear Science · OSTI ID:649407