Vacuum Fizeau wedge wavemeter for pulsed uv lasers

Technical Report ·
OSTI ID:6841578

We have constructed and tested a vacuum-spaced Fizeau wedge wavemeter that is suitable for measuring the wavelength of pulsed or continuous-wave lasers over the range of 630 nm to 200 nm. A Lasertechnics Model 100 Fizeau wavemeter was modified for ultraviolet operation and demonstrated an accuracy of 1.6 parts in 10/sup 5/ at 248 nm. Extensive computer modeling of the wavemeter has provided optimized design parameters for an improved unit. 9 refs., 6 figs., 2 tabs.

Research Organization:
Sandia National Labs., Albuquerque, NM (USA)
DOE Contract Number:
AC04-76DP00789
OSTI ID:
6841578
Report Number(s):
SAND-88-1906; ON: DE88015832
Country of Publication:
United States
Language:
English