Laser-linewidth measurement with a Fizeau wavemeter
Journal Article
·
· Opt. Lett.; (United States)
It is shown theoretically and experimentally that a Fizeau wedge wavemeter typically used to measure laser wavelength can also be used to measure the linewidth of a pulsed or cw laser. Linewidths as small as a few hundred megahertz can be measured by choosing the thickness of the wedge properly.
- Research Organization:
- Sandia National Laboratories, Albuquerque, New Mexico 87185
- OSTI ID:
- 6742511
- Journal Information:
- Opt. Lett.; (United States), Journal Name: Opt. Lett.; (United States) Vol. 13:11; ISSN OPLED
- Country of Publication:
- United States
- Language:
- English
Similar Records
Laser wavemeter with solid Fizeau wedge interferometer
Vacuum Fizeau wedge wavemeter for pulsed uv lasers
Fizeau wavemeter