Field compatible calibration procedure for peroxyacetyl nitrate

Journal Article · · Environ. Sci. Technol.; (United States)
DOI:https://doi.org/10.1021/es00120a013· OSTI ID:6810901
A field compatible calibration procedure for generating ppb levels of peroxyacetyl nitrate is presented. The procedure is based upon the condensed-phase synthesis of peroxyacetyl nitrate (PAN) in octane solution. These solutions show PAN loss of <4% month/sup -1/ when stored at -20/sup 0/C. The dilute PAN solutions are injected directly into known volumes of air in Tedlar bags to obtain ppb gas-phase concentrations. The addition of low levels of NO/sub 2/ into the bags stabilizes PAN in the gas phase.