Skip to main content
U.S. Department of Energy
Office of Scientific and Technical Information

Thermostable cocktail for the liquid scintillation counting of heterogeneous media

Journal Article · · Anal. Biochem.; (United States)
Considerable analytical errors arise in the liquid scintillation counting of heterogeneous media as a consequence of gel instability. With large sample numbers, a major causative factor of this instability is temperature changes during the counting period. An emulsifier-scintillation cocktail has been designed to provide stable counting conditions for heterogeneous media over a temperature range of 10-30/sup 0/C, i.e., the wide range of temperature likely to be encountered in liquid scintillation counters lacking sample cooling facilities. A comparison was made with a conventional commercially available emulsifier-scintillator.
OSTI ID:
6804624
Journal Information:
Anal. Biochem.; (United States), Journal Name: Anal. Biochem.; (United States) Vol. 122:1; ISSN ANBCA
Country of Publication:
United States
Language:
English